Journal of Prosthetic Dentistry

Accuracy and precision of as-received implant torque wrenches

      Statement of problem

      Previous implant torque evaluation did not determine if the target value fell within a confidence interval for the population mean of the test groups, disallowing determination of wheter a specific type of wrench met a standardized goal value.

      Purpose

      The purpose of this study was to measure both the accuracy and precision of 2 different configurations (spring style and peak break) of as-received implant torque wrenches and compare the measured values to manufacturer-stated values.

      Material and methods

      Ten wrenches from 4 manufacturers, representing a variety of torque-limiting mechanisms and specificity of use (with either a specific brand or universally with any brand of implant product). Drivers were placed into the wrench, and tightening torque was applied to reach predetermined values using a NIST-calibrated digital torque wrench. Five replications of measurement were made for each wrench and averaged to provide a single value from that instrument. The target torque value for each wrench brand was compared to the 95% confidence interval for the true population mean of measured values to see if it fell within the measured range.

      Results

      Only 1 wrench brand (Nobel Biocare) demonstrated the target torque value falling within the 95% confidence interval for the true population mean. For others, the targeted torque value fell above the 95% confidence interval (Straumann and Imtec) or below (Salvin Torq).

      Conclusions

      Neither type of torque-limiting mechanism nor designation of a wrench to be used as a dedicated brand-only product or to be used as a universal product on many brands affected the ability of a wrench to deliver torque values where the true population mean included the target torque level.
      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Journal of Prosthetic Dentistry
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Jaarda M.J.
        • Razzoog M.E.
        • Gratton D.G.
        Ultimate tensile strength of five interchangeable prosthetic retaining screws.
        Implant Dent. 1996; 5: 16-19
        • Jaarda M.J.
        • Razzoog M.E.
        • Gratton D.G.
        Comparison of “look-alike” implant prosthetic retaining screws.
        J Prosthodont. 1995; 4: 23-27
        • Jaarda M.J.
        • Razzoog M.E.
        • Gratton D.G.
        Geometric comparison of five interchangeable implant prosthetic retaining screws.
        J Prosthet Dent. 1995; 74: 373-379
        • Burguete R.L.
        • Johns R.B.
        • King T.
        • Patterson E.A.
        Tightening characteristics for screwed joints in osseointegrated dental implants.
        J Prosthet Dent. 1994; 71: 592-599
        • McGlumphy E.A.
        Keeping implant screws tight: the solution.
        J Dent Symp. 1993; 1: 20-23
        • Weinberg L.A.
        The biomechanics of force distribution in implant-supported prostheses.
        Int J Oral Maxillofac Implants. 1993; 8: 19-31
        • Jörn D.
        • Kohorst P.
        • Besdo S.
        • Rücker M.
        • Stiesch M.
        • Borchers L.
        Influence of lubricant on screw preload and stresses in a finite element model for a dental implant.
        J Prosthet Dent. 2014; 112: 340-348
        • Breeding L.C.
        • Dixon D.L.
        • Nelson E.W.
        • Tietge J.D.
        Torque required to loosen single-tooth implant abutment screws before and after simulated function.
        Int J Prosthodont. 1993; 6: 435-439
        • Siamos G.
        • Winkler S.
        • Boberick K.G.
        Relationship between implant preload and screw loosening on implant-supported prostheses.
        J Oral Implantol. 2002; 28: 67-73
        • Winkler S.
        • Ring K.
        • Ring J.D.
        • Boberick K.G.
        Implant screw mechanics and the settling effect: overview.
        J Oral Implantol. 2003; 29: 242-245
        • Sakaguchi R.L.
        • Borgersen S.E.
        Nonlinear contact analysis of preload in dental implant screws.
        Int J Oral Maxillofac Implants. 1995; 10: 295-302
        • Haack J.E.
        • Sakaguchi R.L.
        • Coffey J.P.
        Elongation and preload stress in dental implant abutment screws.
        Int J Oral Maxillofac Implants. 1995; 10: 529-536
        • Bakaeen L.C.
        • Winkler S.
        • Neff P.A.
        The effect of implant diameter, restoration design, and occlusal table variations on screw loosening of posterior single-tooth implant restorations.
        J Oral Implantol. 2001; 27: 63-72
        • Siamos G.
        • Winkler S.
        • Boberick K.G.
        The relationship between implant preload and screw loosening on implant-supported restorations.
        J Oral Implantol. 2002; 28: 67-73
        • Farina A.P.
        • Spazzin A.O.
        • Xediek Consani R.L.
        • Mesquita M.F.
        Screw joint stability after the application of retorque in implant-supported dentures under simulated masticatory conditions.
        J Prosthet Dent. 2014; 111: 499-504
        • Jaarda M.J.
        • Razzoog M.E.
        • Gratton D.G.
        Providing optimum torque to implant prostheses: a pilot study.
        Implant Dent. 1993; 2: 50-52
        • Dellinges M.A.
        • Tebrock O.C.
        A measurement of torque values obtained with hand-held drivers in a simulated clinical setting.
        J Prosthodont. 1993; 2: 212-214
        • Hill E.E.
        • Phillips S.M.
        • Breeding L.C.
        Implant abutment screw torque generated by general dentists using a hand driver in a limited access space simulating the mouth.
        J Oral Implantol. 2007; 33: 277-279
        • Goheen K.L.
        • Vermilyea S.G.
        • Vossoughi J.
        • Agar J.R.
        Torque generated by handheld screwdrivers and mechanical torquing devices for osseointegrated implants.
        Int J Oral Maxillofac Implants. 1994; 9: 149-155
        • Standlee J.P.
        • Caputo A.A.
        • Chwu M.Y.
        • Sun T.T.
        Accuracy of mechanical torque-limiting devices for implants.
        Int J Oral Maxillofac Implants. 2002; 17: 220-224
        • Vallee M.C.
        • Conrad H.J.
        • Basu S.
        • Seong W.J.
        Accuracy of friction-style and spring-style mechanical torque limiting devices for dental implants.
        J Prosthet Dent. 2008; 100: 86-92
        • Akca K.
        • Cehreli M.C.
        Accuracy of 2 impression techniques for ITI implants.
        Int J Oral Maxillofac Implants. 2004; 19: 517-523
        • Gutierrez J.
        • Nicholls J.
        • Libman W.
        • Butson T.
        Accuracy of the implant torque wrench following time in clinical service.
        Int J Prosthodont. 1997; 17: 562-567
        • Standlee J.P.
        • Caputo A.A.
        Accuracy of an electric torque-limiting device for implants.
        Int J Oral Maxillofac Implants. 1999; 14: 278-281
        • Rosner B.
        Fundamentals of biostatistics.
        6th ed. Thomson, Belmont2005: 193
        • De Feo J.A.
        • Juran J.M.
        Juran’s quality handbook: the complete guide to performance excellence.
        6th ed. McGraw-Hill Professional, New York2010: 643-650