Influence of different materials on retention behavior of CAD-CAM fabricated bar attachments

      Abstract

      Statement of problem

      Loss of retention with implant overdentures is a problem that might be addressed with materials such as a recently introduced biocompatible high-performance polymer (BioHPP). However, in vitro studies investigating this product are lacking.

      Purpose

      The purpose of this in vitro study was to assess the retentive behavior of implant overdentures and the loss of retention from clip wear when used with computer-aided design and computer-aided manufacturing (CAD-CAM) fabricated bar attachments from cobalt-chromium (Co-Cr), zirconia (ZrO2), and BioHPP to identify the optimal material in terms of minimal loss of retention and minimal wear.

      Material and methods

      Three CAD-CAM bar attachments were fabricated from the different materials and cemented to a completely edentulous mandibular model for testing. Twelve plastic clips were allocated to each material and subjected to 1095 dislodgement cycles to simulate a year of clinical use. A universal testing machine was used to measure the retentive force of dislodgement of each clip for each material before and after cycling. An inverted optical measuring microscope was used to quantify the amount of wear on each clip. One-way repeated-measures ANOVA was used to analyze the results among the materials and the decrease in retention force. The paired t test was used to compare clip measurements before and after testing (α=.05).

      Results

      BioHPP bars retained retention significantly better than Co-Cr or ZrO2, with a lower percentage decrease in mean retentive force. Clips used with BioHPP bars also exhibited the least wear. However, Co-Cr and ZrO2 displayed similarly higher initial retentive values in comparison with BioHPP, with Co-Cr displaying the highest final retention.

      Conclusions

      BioHPP appears to be a strong candidate to replace Co-Cr or ZrO2 for fabricating bar attachments as it displayed minimal loss of retention and led to the least wear of the clip, thereby reducing replacement frequency.
      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Journal of Prosthetic Dentistry
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Scherer M.D.
        • McGlumphy E.A.
        • Seghi R.R.
        • Campagni W.V.
        Comparison of retention and stability of implant-retained overdentures based upon implant number and distribution.
        Int J Oral Maxillofac Implants. 2013; 28: 1619-1628
        • Critchlow S.B.
        • Ellis J.S.
        Prognostic indicators for conventional complete denture therapy: a review of the literature.
        J Dent. 2010; 38: 2-9
        • van Kampen F.M.C.
        • Cune M.S.
        • van der Bilt A.
        • Bosman F.
        Retention and postinsertion maintenance of bar-clip, ball and magnet attachments in mandibular overdenture treatment: an in vivo comparison after 3 months of function.
        Clin Oral Impl Res. 2003; 14: 720-726
        • Hemmings K.W.
        • Schmitt A.
        • Zarb G.A.
        Complications and maintenance requirements for fixed prostheses and overdentures in the edentulous mandible: a 5-year report.
        Int J Oral Maxillofac Implants. 1994; 9: 191-196
        • Branchi R.
        • Vangi D.
        • Virga A.
        • Guertin G.
        • Fazi G.
        Resistance to wear of four matrices with ball attachments for implant overdentures: a fatigue study.
        J Prosthodont. 2010; 19: 614-619
        • Stephens G.J.
        • di Vitale N.
        • O'Sullivan E.
        • McDonald A.
        The influence of interimplant divergence on the retention characteristics of locator attachments, a laboratory study.
        J Prosthodont. 2014; 23: 467-475
        • Rutkunas V.
        • Mizutani H.
        • Takahashi H.
        Influence of attachment wear on retention of mandibular overdenture.
        J Oral Rehabil. 2007; 34: 41-51
        • Yilmaz B.
        • Ozkir E.
        • Johnston W.M.
        • McGlumphy E.
        Dislodgement force analysis of an overdenture attachment system.
        J Prosthet Dent. 2020; 123: 291-298
        • Salehi R.
        • Shayegh S.S.
        • Johnston W.M.
        • Hakimaneh S.M.
        Effects of interimplant distance and cyclic dislodgement on retention of locator and ball attachments: An in vitro study.
        J Prosthet Dent. 2019; 122: 550-556
        • Aroso C.
        • Silva A.S.
        • Ustrell R.
        • Mendes J.M.
        • Braga A.C.
        • Berastegui E.
        • et al.
        Effect of abutment angulation in the retention and durability of three overdenture attachment systems: An in vitro study.
        J Adv Prosthodont. 2016; 8: 21-29
        • ELsyad M.A.
        • Emera R.M.
        • Ashmawy T.M.
        Effect of distal implant inclination on dislodging forces of different locator attachments used for mandibular overdentures: an in vitro study.
        J Prosthodont. 2019; 28: e666-e674
        • Bayer S.
        • Komor N.
        • Kramer A.
        • Albrecht D.
        • Mericske-Stern R.
        • Enkling N.
        Retention force of plastic clips on implant bars: a randomized controlled trial.
        Clin Oral Implants Res. 2012; 23: 1377-1384
        • Trakas T.
        • Michalakis K.
        • Kang K.
        • Hirayama H.
        Attachment systems for implant retained overdentures: a literature review.
        Implant Dent. 2006; 15: 24-34
        • Karabuda C.
        • Tosun T.
        • Ermis E.
        • Ozdemir T.
        Comparison of 2 retentive systems for implant-supported overdentures: soft tissue management and evaluation of patient satisfaction.
        J Periodontol. 2002; 73: 1067-1070
        • Setz I.
        • Lee S.H.
        • Engel E.
        Retention of prefabricated attachments for implant stabilized overdentures in the edentulous mandible: an in vitro study.
        J Prosthet Dent. 1998; 80: 323-329
        • Benetti A.R.
        • Peutzfeldt A.
        • Lussi A.
        • Flury S.
        Resin composites: Modulus of elasticity and marginal quality.
        J Dent. 2014; 42: 1185-1192
        • Kim D.
        • Park C.
        • Yi Y.
        • Cho L.
        Comparison of cast Ti-Ni alloy clasp retention with conventional removable partial denture clasps.
        J Prosthet Dent. 2004; 91: 374-382
        • McColl I.R.
        • Ding J.
        • Leen S.B.
        Finite element simulation and experimental validation of fretting wear.
        Wear. 2004; 256: 1114-1127
        • Lemm J.D.
        • Warmuth A.R.
        • Pearson S.R.
        • Shipway P.H.
        The influence of surface hardness on the fretting wear of steel pairs—Its role in debris retention in the contact.
        Tribol Int. 2015; 81: 258-266
        • Bridgeport D.A.
        • Brantley W.A.
        • Herman P.F.
        Cobalt-chromium and nickel-chromium alloys for removable prosthodontics, Part 1: Mechanical properties.
        J Prosthodont. 1993; 2: 144-150
        • Xin X.Z.
        • Chen J.
        • Xiang N.
        • Wei B.
        Surface properties and corrosion behavior of Co-Cr alloy fabricated with selective laser melting technique.
        Cell Biochem Biophys. 2013; 67: 983-990
        • Food and Drug Administration H.H.S.
        Dental devices; dental noble metal alloys and dental base metal alloys; designation of special controls. Final rule.
        Fed Regist. 2004; 69: 51765-51766
        • Bathala L.
        • Majeti V.
        • Rachuri N.
        • Singh N.
        • Gedela S.
        The role of polyether ether ketone (PEEK) in dentistry–a review.
        J Med Life. 2019; 12: 5-9
        • Zoidis P.
        • Papathanasiou I.
        • Polyzois G.
        The use of a modified poly-ether-ether-ketone (PEEK) as an alternative framework material for removable dental prostheses. A clinical report.
        J Prosthodont. 2016; 25: 580-584
        • Sailer I.
        • Feher A.
        • Filser F.
        • Gauckler L.J.
        • Luthy H.
        • Hammerle C.H.
        Five-year clinical results of zirconia frameworks for posterior fixed partial dentures.
        Int J Prosthodont. 2007; 20: 383-388
        • Guazzato M.
        • Albakry M.
        • Swain M.V.
        • Ironside J.
        Mechanical properties of in-Ceram alumina and in-Ceram zirconia.
        Int J Prosthodont. 2002; 15: 339-346
        • Tinschert J.
        • Natt G.
        • Mautsch W.
        • Augthun M.
        • Spiekermann H.
        Fracture resistance of lithium disilicate, alumina, and zirconia-based three-unit fixed partial dentures: a laboratory study.
        Int J Prosthodont. 2001; 14: 231-238
        • Pittayachawan P.
        • McDonald A.
        • Petrie A.
        • Knowles J.C.
        The biaxial flexural strength and fatigue property of Lava Y-TZP dental ceramic.
        Dent Mater. 2007; 23: 1018-1029
        • Lawson N.C.
        • Janyavula S.
        • Syklawer S.
        • McLaren E.A.
        • Burgess J.O.
        Wear of enamel opposing zirconia and lithium disilicate after adjustment, polishing and glazing.
        J Dent. 2014; 42: 1586-1591
        • Park J.H.
        • Park S.
        • Lee K.
        • Yun K.D.
        • Lim H.P.
        Antagonist wear of three CAD/CAM anatomic contour zirconia ceramics.
        J Prosthet Dent. 2014; 111: 20-29
        • Kontos L.
        • Schille C.
        • Schweizer E.
        • Geis-Gerstorfer J.
        Influence of surface treatment on the wear of solid zirconia.
        Acta Odontol Scand. 2013; 71: 482-487
        • Passos S.P.
        • Torrealba Y.
        • Major P.
        • Linke B.
        • Flores-Mir C.
        • Nychka J.A.
        In vitro wear behavior of zirconia opposing enamel: a systematic review.
        J Prosthodont. 2014; 23: 593-601
        • Vagkopoulou T.
        • Koutayas S.O.
        • Koidis P.
        • Strub J.R.
        Zirconia in dentistry: Part 1. Discovering the nature of an upcoming bioceramic.
        Eur J Esthet Dent. 2009; 4: 130-151
        • Kurtz S.M.
        • Devine J.N.
        PEEK biomaterials in trauma, orthopedic, and spinal implants.
        Biomaterials. 2007; 28: 4845-4869
        • Ha S.W.
        • Kirch M.
        • Birchler F.
        • Eckert K.L.
        • Mayer J.
        • Wintermantel E.
        Surface activation of polyetheretherketone (PEEK) and formation of calcium phosphate coatings by precipitation.
        J Mater Sci Mater Med. 1997; 8: 683-690
        • Goyal R.
        • Tiwari A.
        • Negi Y.
        Microhardness of PEEK/ceramic micro-and nanocomposites: correlation with Halpin–Tsai model.
        Mater Sci Eng A. 2008; 491: 230-236
        • Wang L.
        • Weng L.
        • Wu Z.
        • Wang C.
        The properties of polyetheretherketone biocomposite reinforced by surface-modified nano-hydroxyapatite.
        Adv Mat Res. 2015; 1096: 214-218
        • Kassem Y.
        • Alshimy A.M.
        • El-Shrabawy S.M.
        Mechanical evaluation of polyetheretherketone compared with zirconia as a dental implant material.
        ADJ. 2019; 44: 61-66
        • Menicucci G.
        • Lorenzetti M.
        • Pera P.
        • Preti G.
        Mandibular implant-retained overdenture: finite element analysis of two anchorage systems.
        Int J Oral Maxillofac Implants. 1998; 13: 369-376
        • Pasciuta M.
        • Grossmann Y.
        • Finger I.M.
        A prosthetic solution to restoring the edentulous mandible with limited interarch space using an implant-tissue-supported overdenture: a clinical report.
        J Prosthet Dent. 2005; 93: 116-120
        • Sinclair P.M.
        • Little R.M.
        Maturation of untreated normal occlusions.
        Am J Orthod. 1983; 83: 114-123
        • Choi J.W.
        • Bae J.H.
        • Jeong C.M.
        • Huh J.B.
        Retention and wear behaviors of two implant overdenture stud-type attachments at different implant angulations.
        J Prosthet Dent. 2017; 117: 628-635
        • Alao A.R.
        • Stoll R.
        • Song X.F.
        • Miyazaki T.
        • Hotta Y.
        • Shibata Y.
        • et al.
        Surface quality of yttria-stabilized tetragonal zirconia polycrystal in CAD/CAM milling, sintering, polishing and sandblasting processes.
        J Mech Behav Biomed. 2017; 65: 102-116
        • Barbour M.E.
        • O’Sullivan D.J.
        • Jenkinson H.F.
        • Jagger D.C.
        The effects of polishing methods on surface morphology, roughness and bacterial colonisation of titanium abutments.
        J Mater Sci Mater Med. 2007; 18: 1439-1447
        • Ozcelik T.B.
        • Ozcelik O.
        • Pektas Z.O.
        Cement-retained bar attachment system for implant-retained mandibular overdentures.
        J Prosthet Dent. 2007; 97: 315-317
        • Katsoulis J.
        • Mericske-Stern R.
        • Yates D.M.
        • Izutani N.
        • Enkling N.
        • Blatz M.B.
        In vitro precision of fit of computer-aided design and computer-aided manufacturing titanium and zirconium dioxide bars.
        Dent Mater. 2013; 29: 945-953
        • Chung K.H.
        • Chung C.Y.
        • Cagna D.R.
        • Cronin Jr., R.J.
        Retention characteristics of attachment systems for implant overdentures.
        J Prosthodont. 2004; 13: 221-226
        • Besimo C.E.
        • Guarneri A.
        In vitro retention force changes of prefabricated attachments for overdentures.
        J Oral Rehabil. 2003; 30: 671-678
        • Dean A.G.
        • Dean J.A.
        • Burton A.H.
        • Dicker R.C.
        Epi Info: a general-purpose microcomputer program for public health information systems.
        Am. J. Prev. Med. 1991; 7: 178-182
        • McCrum-Gardner E.
        Sample size and power calculations made simple.
        Int J Rehabil Res. 2010; 17: 10-14
        • Ozkir S.E.
        • Unal S.M.
        • Yurekli E.
        • Güven S.
        Effects of crown retrieval on implants and the surrounding bone: a finite element analysis.
        J Adv Prosthodont. 2016; 8: 131-136
        • Kahneman D.
        • Knetsch J.L.
        • Thaler R.H.
        Anomalies: the endowment effect, loss aversion, and status quo bias.
        J Econ Perspect. 1991; 5: 193-206