Influence of connector cross-sectional geometry on the load-bearing capacity under fatigue of implant-supported zirconia fixed partial prosthesis

Published:November 17, 2022DOI:https://doi.org/10.1016/j.prosdent.2022.10.008

      Abstract

      Statement of problem

      Zirconia has been reported to be suitable for multiunit monolithic restorations. However, as the connector region is the weakest part of the system, studies are needed to determine the optimal connector geometry.

      Purpose

      The purpose of this in vitro study was to evaluate the load-bearing capacity under fatigue of implant-supported fixed partial prostheses made of monolithic zirconia with different connector cross-sectional geometries.

      Material and methods

      Three-unit monolithic fixed partial prostheses (from mandibular second premolar to the second molar) were fabricated in zirconia (Zenostar T) by computer-aided design and computer-aided manufacture (CAD-CAM). Different connector cross-sectional geometries, with an area of 9 mm2, were tested (n=10): round, square with rounded angles, or trapezoid with rounded angles. The prostheses were screwed over 2 implants and inserted into acrylic resin bases. The specimens were submitted to a mechanical fatigue test until failure, with load applied to the pontic using the stepwise stress methodology (initial load of 100 N for 5000 cycles, followed by an increase to 200 N for 20 000 cycles, with a subsequent increase of 50 N each step). For data analysis, fatigue failure load (FFL) and cycles for fatigue failure (CFF) were recorded. Failed structures were analyzed by fractography. The Kaplan-Meier test followed by a log-rank Mantel-Cox post hoc test was used to analyze FFL and CFF (α=.05). Weibull analysis was also performed.

      Results

      The round (1065 N) and trapezoid (1010 N) groups presented higher FFL values than the square one (870 N) (P≤.05). For CFF, the round connector (358 777) was better than the square (280 310) (P=.006), and the trapezoid (337 773) was statistically similar to both (P>.05). No difference among groups was observed in Weibull modulus either for CFF or FFL data. All failures were catastrophic, originating at the base of the connector and propagating toward the occlusal surface of the pontic (region under loading).

      Conclusions

      The connector cross-sectional geometry significantly influenced the mechanical fatigue performance of implant-supported fixed partial prostheses made of monolithic zirconia.
      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Journal of Prosthetic Dentistry
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Sailer I.
        • Strasding M.
        • Valente N.A.
        • Zwahlen M.
        • Liu S.
        • Pjetursson B.E.
        A systematic review of the survival and complication rates of zirconia-ceramic and metal-ceramic multiple-unit fixed dental prostheses.
        Clin Oral Implants Res. 2018; 16: 184-198
        • Bacchi A.
        • Cesar P.F.
        Advances in ceramics for dental applications.
        Dent Clin North Am. 2022; 66: 591-602
        • Pjetursson B.E.
        • Valente N.A.
        • Strasding M.
        • Zwahlen M.
        • Liu S.
        • Sailer I.
        A systematic review of the survival and complication rates of zirconia-ceramic and metal-ceramic single crowns.
        Clin Oral Implants Res. 2018; 16: 199-214
        • Morton D.
        • Gallucci G.
        • Lin W.S.
        • et al.
        Group 2 ITI consensus report: prosthodontics and implant dentistry.
        Clin Oral Implants Res. 2018; 16: 215-223
        • Zhang Y.
        • Lawn B.R.
        Novel zirconia materials in dentistry.
        J Dent Res. 2018; 97: 140-147
        • Stawarczyk B.
        • Keul C.
        • Eichberger M.
        • Figge D.
        • Edelhoff D.
        • Lumkemann N.
        Three generations of zirconia: from veneered to monolithic. Part I.
        Quintessence Int. 2017; 48: 369-380
        • Stawarczyk B.
        • Keul C.
        • Eichberger M.
        • Figge D.
        • Edelhoff D.
        • Lumkemann N.
        Three generations of zirconia: from veneered to monolithic. Part II.
        Quintessence Int. 2017; 48: 441-450
        • Pereira G.K.R.
        • Guilardi L.F.
        • Dapieve K.S.
        • Kleverlaan C.J.
        • Rippe M.P.
        • Valandro L.F.
        Mechanical reliability, fatigue strength and survival analysis of new polycrystalline translucent zirconia ceramics for monolithic restorations.
        J Mech Behav Biomed Mater. 2018; 85: 57-65
        • Kolakarnprasert N.
        • Kaizer M.R.
        • Kim D.K.
        • Zhang Y.
        New multi-layered zirconias: composition, microstructure and translucency.
        Dent Mater. 2019; 35: 797-806
        • Brizuela-Velasco A.
        • Diéguez-Pereira M.
        • Álvarez-Arenal Á.
        • et al.
        Fracture resistance of monolithic high translucency zirconia implant-supported crowns.
        Implant Dent. 2016; 25: 624-628
        • Vafaee F.
        • Firouz F.
        • Khoshhal M.
        • Hooshyarfard A.
        • Shahbazi A.
        • Roshanaei G.
        Fatigue fracture strength of implant-supported full contour zirconia and metal ceramic fixed partial dentures.
        J Dent (Tehran). 2017; 14: 165-172
        • Zhang Y.
        • Sailer I.
        • Lawn B.R.
        Fatigue of dental ceramics.
        J Dent. 2013; 41: 1135-1147
        • Oh W.S.
        • Anusavice K.J.
        Effect of connector design on the fracture resistance of all-ceramic fixed partial dentures.
        J Prosthet Dent. 2002; 87: 536-542
        • Larsson C.
        • Holm L.
        • Lövgren N.
        • Kokubo Y.
        • Vult von Steyern P.
        Fracture strength of four-unit Y-TZP FPD cores designed with varying connector diameter. An in-vitro study.
        J Oral Rehabil. 2007; 34: 702-709
        • Plengsombut K.
        • Brewer J.D.
        • Monaco E.A.J.
        • Davis E.L.
        Effect of two connector designs on the fracture resistance of all-ceramic core materials for fixed dental prostheses.
        J Prosthet Dent. 2009; 101: 166-173
        • Quinn G.D.
        • Studart A.R.
        • Hebert C.
        • VerHoef J.R.
        • Arola D.
        Fatigue of zirconia and dental bridge geometry: design implications.
        Dent Mater. 2010; 26: 1133-1136
        • Möllers K.
        • Pätzold W.
        • Parkot D.
        • et al.
        Influence of connector design and material composition and veneering on the stress distribution of all-ceramic fixed dental prostheses: a finite element study.
        Dent Mater. 2011; 27: e171-e175
        • Murase T.
        • Nomoto S.
        • Sato T.
        • Shinya A.
        • Koshihara T.
        • Yasuda H.
        Effect of connector design on fracture resistance in all-ceramic fixed partial dentures for mandibular incisor region.
        Bull Tokyo Dent Coll. 2014; 55: 149-155
        • Hamza T.A.
        • Attia M.A.
        • El-Hossary M.M.K.
        • Mosleh I.E.
        • Shokry T.E.
        • Wee A.G.
        Flexural strength of small connector designs of zirconia-based partial fixed dental prostheses.
        J Prosthet Dent. 2016; 115: 224-229
        • Heintze S.D.
        • Monreal D.
        • Reinhardt M.
        • et al.
        Fatigue resistance of all-ceramic fixed partial dentures - fatigue tests and finite element analysis.
        Dent Mater. 2018; 34: 494-507
        • Bakitian F.
        • Seweryniak P.
        • Papia E.
        • Larsson C.
        • Vult von Steyern P.
        Load-bearing capacity of monolithic zirconia fixed dental prostheses fabricated with different connector designs and embrasure shaping methods.
        J Prosthodont. 2019; 28: 64-70
        • SaranBabu K.A.
        • Perisetty D.K.
        • Thota G.
        • Rasool M.
        • Niharika M.
        • Swapna S.
        Influence of radius of curvature at gingival embrasure in connector area on stress distribution of three-unit posterior full-contour monolithic zirconia fixed partial denture on various amounts of load application: a finite element study.
        J Int Soc Prev Community Dent. 2019; 9: 338-348
        • Ravidà A.
        • Tattan M.
        • Askar H.
        • Barootchi S.
        • Tavelli L.
        • Wang H.L.
        Comparison of three different types of implant-supported fixed dental prostheses: a long-term retrospective study of clinical outcomes and cost-effectiveness.
        Clin Oral Implants Res. 2019; 30: 295-305
        • Worni A.
        • Kolgeci L.
        • Rentsch-Kollar A.
        • Katsoulis J.
        • Mericske-Stern R.
        Zirconia-based screw-retained prostheses supported by implants: a retrospective study on technical complications and failures.
        Clin Implant Dent Relat Res. 2015; 17: 1073-1081
        • Magne P.
        • Carvalho A.O.
        • Bruzi G.
        • Giannini M.
        Fatigue resistance of ultrathin CAD/CAM complete crowns with a simplified cementation process.
        J Prosthet Dent. 2015; 114: 574-579
        • Kelly J.R.
        Clinically relevant approach to failure testing of all-ceramic restorations.
        J Prosthet Dent. 1999; 81: 652-661
        • Fraga S.
        • Pereira G.K.R.
        • Freitas M.
        • Kleverlaan C.J.
        • Valandro L.F.
        • May L.G.
        Loading frequencies up to 20 Hz as an alternative to accelerate fatigue strength tests in a Y-TZP ceramic.
        J Mech Behav Biomed Mater. 2016; 61: 79-86
        • Kelly J.R.
        • Campbell S.D.
        • Bowen H.K.
        Fracture-surface analysis of dental ceramics.
        J Prosthet Dent. 1989; 62: 536-541
        • Yen T.W.
        • Blackman R.B.
        • Baez R.J.
        Effect of acid etching on the flexural strength of a feldspathic porcelain and a castable glass ceramic.
        J Prosthet Dent. 1993; 70: 224-233
        • Thompson J.Y.
        • Anusavice K.J.
        • Naman A.
        • Morris H.F.
        Fracture surface characterization of clinically failed all-ceramic crowns.
        J Dent Res. 1994; 73: 1824-1832
        • Thompson J.Y.
        • Anusavice K.J.
        Effect of surface etching on the flexure strength and fracture toughness of Dicor disks containing controlled flaws.
        J Dent Res. 1994; 73: 505-510
        • Nishigawa K.
        • Bando E.
        • Nakano M.
        Quantitative study of bite force during sleep associated bruxism.
        J Oral Rehabil. 2001; 28: 485-491