Trueness of crowns fabricated by using additively and subtractively manufactured resin-based CAD-CAM materials

  • Gülce Çakmak
    Affiliations
    Senior Research Associate, Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland
    Search for articles by this author
  • Ana Maria Rusa
    Affiliations
    Dissertation Student, Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland
    Search for articles by this author
  • Mustafa Borga Donmez
    Correspondence
    Corresponding author: Dr Mustafa Borga Dönmez, Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Freiburgstrasse 7, Bern 3007, SWITZERLAND
    Affiliations
    Asisstant Professor, Department of Prosthodontics, Istinye University, Faculty of Dentistry, İstanbul, Turkey

    Visiting Researcher, Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland
    Search for articles by this author
  • Canan Akay
    Affiliations
    Associate Professor, Department of Prosthodontics, Faculty of Dentistry, Osmangazi University, Eskisehir, Turkey

    Associate Professor, Translational Medicine Research and Clinical Center, Osmangazi University, Eskisehir, Turkey
    Search for articles by this author
  • Çiğdem Kahveci
    Affiliations
    Assistant Professor, Department of Prosthodontics, Giresun University Faculty of Dentistry, Giresun, Turkey
    Search for articles by this author
  • Martin Schimmel
    Affiliations
    Chairman, Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland

    External Research Associate, Division of Gerodontology and Removable Prosthodontics, University Clinics of Dental Medicine, University of Geneva, Geneva, Switzerland
    Search for articles by this author
  • Burak Yilmaz
    Affiliations
    Associate Professor, Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland

    Associate Professor, Department of Restorative, Preventive and Pediatric Dentistry, School of Dental Medicine, University of Bern, Bern, Switzerland

    Adjunct Professor, Division of Restorative and Prosthetic Dentistry, The Ohio State University College of Dentistry, Columbus, Ohio
    Search for articles by this author
Open AccessPublished:December 02, 2022DOI:https://doi.org/10.1016/j.prosdent.2022.10.012

      Abstract

      Statement of problem

      Advancements in digital dental technologies have enabled the use of different resin-based materials that can be fabricated either additively or subtractively. However, knowledge on the fabrication trueness of these materials is scarce.

      Purpose

      The purpose of this in vitro study was to investigate the trueness of crowns fabricated by using different resin-based computer-aided design and computer-aided manufacturing (CAD-CAM) materials.

      Material and methods

      A complete crown for a mandibular right first molar with a 30-μm cement space was designed in standard tessellation language (STL) format. This master STL (MC-STL) was used to fabricate 40 complete crowns with 4 different resin-based CAD-CAM materials and either additive (Crowntec [MS]) or subtractive techniques (Brilliant Crios [BC], breCAM.monoCOM [PMMA], and G-CAM [GR]; n=10). All crowns were digitized with an intraoral scanner (CEREC Primescan SW 5.2) to generate their STL files (TC-STLs). MC-STL and TC-STLs were transferred into a 3-dimensional analysis software program (Medit Link v2.4.4), and a trueness (overall, external, occlusal, intaglio occlusal, and marginal) analysis was performed by using the root mean square (RMS) method. The Kruskal-Wallis and Dunn tests were performed to analyze data (α=.05).

      Results

      The test groups had significantly different deviations on all surfaces (P≤.001). MS crowns had higher overall (P≤.007) and external surface (P≤.001) deviations than GR and PMMA crowns, while the differences between GR and PMMA crowns were not significant (P≥.441). BC crowns had higher external surface deviations than GR crowns (P=.005), higher occlusal deviations than GR and MS crowns (P≤.007), and higher intaglio occlusal deviations than GR and MS crowns (P≤.009). However, BC crowns had lower marginal deviations than MS and GR crowns (P≤.018).

      Conclusions

      The brand of resin-based CAD-CAM materials affected the trueness of crowns. Additively manufactured crowns (MS) mostly had lower overall and external surface trueness than the other groups. Nevertheless, the deviation values of occlusal, intaglio occlusal, and marginal trueness were generally small; thus, the effect of the tested materials on clinical crown fit may be negligible.
      Clinical Implications
      The clinical fit of complete crowns fabricated by using the tested materials was similar. However, crowns fabricated by using the tested additively manufactured resin might require more chairside adjustments.
      Digital technologies have revolutionized prosthodontics,
      • Revilla-León M.
      • Jiang P.
      • Sadeghpour M.
      • et al.
      Intraoral digital scans-part 1: influence of ambient scanning light conditions on the accuracy (trueness and precision) of different intraoral scanners.
      • Revilla-León M.
      • Jiang P.
      • Sadeghpour M.
      • et al.
      Intraoral digital scans: part 2-influence of ambient scanning light conditions on the mesh quality of different intraoral scanners.
      • Cakmak G.
      • Marques V.R.
      • Donmez M.B.
      • Lu W.E.
      • Abou-Ayash S.
      • Yilmaz B.
      Comparison of measured deviations in digital implant scans depending on software and operator.
      and computer-aided design and computer-aided manufacturing (CAD-CAM) systems have been integrated into daily practice by means of subtractive manufacturing
      • Beuer F.
      • Schweiger J.
      • Edelhoff D.
      Digital dentistry: an overview of recent developments for CAD/CAM generated restorations.
      ,
      • Son K.
      • Lee J.H.
      • Lee K.B.
      Comparison of intaglio surface trueness of interim dental crowns fabricated with SLA 3D printing, DLP 3D printing, and milling technologies.
      and, more recently, additive manufacturing.
      • Moon W.
      • Kim S.
      • Lim B.S.
      • Park Y.S.
      • Kim R.J.
      • Chung S.H.
      Dimensional accuracy evaluation of temporary dental restorations with different 3D printing systems.
      ,
      • Alharbi N.
      • Osman R.B.
      • Wismeijer D.
      Factors ınfluencing the dimensional accuracy of 3D-printed full-coverage dental restorations using stereolithography technology.
      Of many available additive manufacturing technologies,
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness of 3-unit interim fixed partial dentures.
      digital light processing (DLP) has been commonly used for dental procedures.
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness and margin quality of 3D-printed ınterim dental crowns.
      ,
      • Çakmak G.
      • Donmez M.B.
      • Cuellar A.R.
      • Kahveci Ç.
      • Schimmel M.
      • Yilmaz B.
      Additive or subtractive manufacturing of crown patterns used for pressing or casting: a trueness analysis.
      Regardless of the technology, additive manufacturing improved the manufacturing processes by enabling cost-efficient fabrication with less waste and the fabrication of products with more complex geometries.
      • Shim J.S.
      • Kim J.E.
      • Jeong S.H.
      • Choi Y.J.
      • Ryu J.J.
      Printing accuracy, mechanical properties, surface characteristics, and microbial adhesion of 3D-printed resins with various printing orientations.
      • Yu B.Y.
      • Son K.
      • Lee K.B.
      Evaluation of intaglio surface trueness and margin quality of interim crowns in accordance with the build angle of stereolithography apparatus 3-dimensional printing.
      • Lerner H.
      • Nagy K.
      • Pranno N.
      • Zarone F.
      • Admakin O.
      • Mangano F.
      Trueness and precision of 3D-printed versus milled monolithic zirconia crowns: an in vitro study.
      Advancements in CAD-CAM technologies have also diversified the materials that can be used either with subtractive manufacturing or additive manufacturing.
      • Peng C.C.
      • Chung K.H.
      • Yau H.T.
      • Ramos Jr., V.
      Assessment of the internal fit and marginal integrity of interim crowns made by different manufacturing methods.
      • Guo N.
      • Leu M.C.
      Additive manufacturing: technology, applications and research needs.
      • Ciocan L.T.
      • Ghitman J.
      • Vasilescu V.G.
      • Iovu H.
      Mechanical properties of polymer-based blanks for machined dental restorations.
      • Corbani K.
      • Hardan L.
      • Eid R.
      • et al.
      Fracture resistance of three-unit fixed dental prostheses fabricated with milled and 3D printed composite-based materials.
      One of the recent advancements in subtractively manufactured restorative materials was the integration of graphene as a reinforcement phase in polymers, including polymethyl methacrylate (PMMA), through nanotechnology.
      • Agarwalla S.V.
      • Malhotra R.
      • Rosa V.
      Translucency, hardness and strength parameters of PMMA resin containing graphene-like material for CAD/CAM restorations.
      Graphene is a crystalline form of carbon
      • Di Carlo S.
      • De Angelis F.
      • Brauner E.
      • et al.
      Flexural strength and elastic modulus evaluation of structures made by conventional PMMA and PMMA reinforced with graphene.
      that has favorable mechanical properties
      • Hernández J.
      • Mora K.
      • Boquete-Castro A.
      • Kina S.
      The effect of thermocycling on surface microhardness of PMMA doped with graphene: an experimental in vitro study.
      which have been reported to improve the properties of PMMA,
      • Ciocan L.T.
      • Ghitman J.
      • Vasilescu V.G.
      • Iovu H.
      Mechanical properties of polymer-based blanks for machined dental restorations.
      ,
      • Di Carlo S.
      • De Angelis F.
      • Brauner E.
      • et al.
      Flexural strength and elastic modulus evaluation of structures made by conventional PMMA and PMMA reinforced with graphene.
      ,
      • Lee J.-H.
      • Jo J.-K.
      • Kim D.-A.
      • Patel K.D.
      • Kim H.-W.
      • Lee H.-H.
      Nano-graphene oxide incorporated into PMMA resin to prevent microbial adhesion.
      and a nanographene-reinforced PMMA, indicated for veneers, inlays, onlays, tooth- or implant-supported crowns, and 3-unit tooth-supported fixed partial dentures, has been marketed. The additive manufacturing of dental products has become more popular given the advantages of this technology over subtractive manufacturing,
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness of 3-unit interim fixed partial dentures.
      and additively manufactured composite resins that can be used for definitive prostheses have been recently introduced.
      • Grzebieluch W.
      • Kowalewski P.
      • Grygier D.
      • Rutkowska-Gorczyca M.
      • Kozakiewicz M.
      • Jurczyszyn K.
      Printable and machinable dental restorative composites for CAD/CAM application—comparison of mechanical properties, fractographic, texture and fractal dimension analysis.
      Even though these materials have been indicated for use in definitive prostheses by their manufacturers, the properties of prostheses fabricated by using materials that could affect their clinical success should be broadly investigated. The studies focusing on different properties of nanographene-reinforced PMMA
      • Ciocan L.T.
      • Ghitman J.
      • Vasilescu V.G.
      • Iovu H.
      Mechanical properties of polymer-based blanks for machined dental restorations.
      ,
      • Agarwalla S.V.
      • Malhotra R.
      • Rosa V.
      Translucency, hardness and strength parameters of PMMA resin containing graphene-like material for CAD/CAM restorations.
      • Di Carlo S.
      • De Angelis F.
      • Brauner E.
      • et al.
      Flexural strength and elastic modulus evaluation of structures made by conventional PMMA and PMMA reinforced with graphene.
      • Hernández J.
      • Mora K.
      • Boquete-Castro A.
      • Kina S.
      The effect of thermocycling on surface microhardness of PMMA doped with graphene: an experimental in vitro study.
      • Lee J.-H.
      • Jo J.-K.
      • Kim D.-A.
      • Patel K.D.
      • Kim H.-W.
      • Lee H.-H.
      Nano-graphene oxide incorporated into PMMA resin to prevent microbial adhesion.
      or additively manufactured composite resins
      • Corbani K.
      • Hardan L.
      • Eid R.
      • et al.
      Fracture resistance of three-unit fixed dental prostheses fabricated with milled and 3D printed composite-based materials.
      ,
      • Corbani K.
      • Hardan L.
      • Skienhe H.
      • Özcan M.
      • Alharbi N.
      • Salameh Z.
      Effect of material thickness on the fracture resistance and failure pattern of 3D-printed composite crowns.
      • Zimmermann M.
      • Ender A.
      • Attin T.
      • Mehl A.
      Fracture load of three-unit full-contour fixed dental prostheses fabricated with subtractive and additive CAD/CAM technology.
      • Zimmermann M.
      • Ender A.
      • Egli G.
      • Özcan M.
      • Mehl A.
      Fracture load of CAD/CAM-fabricated and 3D-printed composite crowns as a function of material thickness.
      • Atria P.J.
      • Bordin D.
      • Marti F.
      • et al.
      3D-printed resins for provisional dental restorations: comparison of mechanical and biological properties.
      • Donmez M.B.
      • Okutan Y.
      Marginal gap and fracture resistance of implant-supported 3D-printed definitive composite crowns: an in vitro study.
      were not based on the trueness of the manufactured products. Considering that the optimal fit of a prosthesis is correlated with its dimensional accuracy,
      • Pacquet W.
      • Tapie L.
      • Mawussi B.
      • Boitelle P.
      Volumetric and dimensional accuracy assessment of CAD-CAM–manufactured dental prostheses from different materials.
      studies focusing on the trueness of prostheses fabricated with different types of CAD-CAM materials could benefit clinicians. Advancements in digital technologies have also facilitated these analyses, and it is possible to compare a product with its CAD file.
      • Cakmak G.
      • Marques V.R.
      • Donmez M.B.
      • Lu W.E.
      • Abou-Ayash S.
      • Yilmaz B.
      Comparison of measured deviations in digital implant scans depending on software and operator.
      Therefore, the aim of the present study was to evaluate the trueness of CAD-CAM crowns made of 4 different resin-based materials that are either additively (1 additively manufactured composite resin) or subtractively manufactured (1 reinforced composite resin, 1 PMMA, and 1 nanographene-reinforced PMMA) by using digital comparison tools. The null hypothesis was that material type would not affect the trueness of crowns fabricated by using different CAD-CAM technologies.

      Material and methods

      Figure 1 illustrates the methodology of the present study. A prefabricated titanium abutment with a 1.5-mm-wide chamfer finish line was digitized by using an intraoral scanner (CEREC Primescan SW 5.2; Dentsply Sirona). A complete crown for a mandibular right first molar with a 30-μm cement gap,
      • Alharbi N.
      • Osman R.B.
      • Wismeijer D.
      Factors ınfluencing the dimensional accuracy of 3D-printed full-coverage dental restorations using stereolithography technology.
      ,
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness and margin quality of 3D-printed ınterim dental crowns.
      ,
      • Çakmak G.
      • Donmez M.B.
      • Cuellar A.R.
      • Kahveci Ç.
      • Schimmel M.
      • Yilmaz B.
      Additive or subtractive manufacturing of crown patterns used for pressing or casting: a trueness analysis.
      3-mm-thick axial walls, 1.5-mm-thick margins, and 1 mm of minimum occlusal thickness was designed on the standard tessellation language (STL) file of the abutment by using a dental design software program (exocad DentalCAD; exocad GmbH).
      Figure thumbnail gr1
      Figure 1Overview of study. CAD-CAM, computer-aided design and computer-aided manufacturing. BC, Brilliant Crios; CAD-CAM, computer-aided design and computer-aided manufacturing; GR, G-CAM; MS, Crowntec; PMMA, breCAM.monoCOM.
      The number of specimens in each group was determined based on previous studies investigating the trueness of additively manufactured prostheses that reported significant differences.
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness of 3-unit interim fixed partial dentures.
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness and margin quality of 3D-printed ınterim dental crowns.
      • Çakmak G.
      • Donmez M.B.
      • Cuellar A.R.
      • Kahveci Ç.
      • Schimmel M.
      • Yilmaz B.
      Additive or subtractive manufacturing of crown patterns used for pressing or casting: a trueness analysis.
      ,
      • Lerner H.
      • Nagy K.
      • Pranno N.
      • Zarone F.
      • Admakin O.
      • Mangano F.
      Trueness and precision of 3D-printed versus milled monolithic zirconia crowns: an in vitro study.
      This master crown STL (MC-STL) was used to fabricate 40 complete-coverage crowns by using 4 different CAD-CAM materials (Brilliant Crios; Coltène AG [BC], breCAM.monoCOM; bredent [PMMA], Crowntec; Saremco Dental AG [MS], and G-CAM; Graphenano DENTAL SL [GR]). For the fabrication of subtractively manufactured crowns (BC, PMMA, and GR), MC-STL was transferred into a nesting software program (PrograMill CAM V4; Ivoclar AG) and milled with a 5-axis milling unit (PrograMill PM7; Ivoclar AG) (n=10).
      For the fabrication of additively manufactured crowns (MS), MC-STL was imported into a nesting software program (Composer; ASIGA) and positioned with their occlusal surface facing toward the build platform. Support structures were generated automatically, and any support on the margin or in the intaglio surface was eliminated manually. This configuration was duplicated 10 times, and the layer thickness was set at 50 μm according to the manufacturer’s recommendations.
      • Donmez M.B.
      • Okutan Y.
      Marginal gap and fracture resistance of implant-supported 3D-printed definitive composite crowns: an in vitro study.
      MS crowns were printed by using the proprietary DLP-based 3-dimensional (3D) printer (MAX UV; ASIGA). After printing, the crowns were removed from the build platform, cleaned with an alcohol-soaked (96% Alcohol isopropilico; Quinei Klean) cloth until all unpolymerized resin had been removed, and dried with an air syringe. Crowns were then placed in a xenon polymerization device (Otoflash G171; NK Optik) in an atmosphere of nitrogen oxide gas for 4000 light exposures.
      • Donmez M.B.
      • Okutan Y.
      Marginal gap and fracture resistance of implant-supported 3D-printed definitive composite crowns: an in vitro study.
      After the fabrication of additively and subtractively manufactured crowns had been completed, the support structures were removed with a cut-off-wheel (Keystone Cut-off Wheels; Keystone Industries), and the surface was gently smoothed under optical magnification loupes (EyeMag Pro; Carl Zeiss) at ×3.5 magnification to prevent errors during the alignment procedure. No adjustments were made on the intaglio surfaces (Fig. 2).
      • Alharbi N.
      • Osman R.B.
      • Wismeijer D.
      Factors ınfluencing the dimensional accuracy of 3D-printed full-coverage dental restorations using stereolithography technology.
      Figure thumbnail gr2
      Figure 2CAD-CAM fabricated crowns. A, Occlusal surface. B, Buccal surface. C, Intaglio surface and margin. D, Buccal view on die. E, Lingual view on die. CAD-CAM, computer-aided design and computer-aided manufacturing. BC, Brilliant Crios; CAD-CAM, computer-aided design and computer-aided manufacturing; GR, G-CAM; MS, Crowntec; PMMA, breCAM.monoCOM.
      A single operator (A.M.R.) digitized all crowns by using the same intraoral scanner to generate the STL files of the crowns (TC-STLs). The intraoral scanner was calibrated before starting the scan of each group, and fatigue-related deviations were minimized as the operator took 5-minute breaks between each group.
      • Mangano F.
      • Lerner H.
      • Margiani B.
      • Solop I.
      • Latuta N.
      • Admakin O.
      Congruence between meshes and library files of ımplant scanbodies: an ın vitro study comparing five ıntraoral scanners.
      All scans were performed in the same temperature and humidity-controlled room as the scan of the titanium abutment.
      All STL files (1 MC-STL and 40 TC-STLs) were imported into a 3D analysis software program (Medit Link v2.4.4; Medit).
      • Cakmak G.
      • Marques V.R.
      • Donmez M.B.
      • Lu W.E.
      • Abou-Ayash S.
      • Yilmaz B.
      Comparison of measured deviations in digital implant scans depending on software and operator.
      ,
      • Yilmaz B.
      • Marques V.R.
      • Donmez M.B.
      • et al.
      Influence of 3D analysis software on measured deviations of CAD-CAM resin crowns from virtual design file: an in-vitro study.
      MC-STL was selected as the reference, and the comparison tool of the software program was used to superimpose TC-STL over the MC-STL by simultaneously selecting 3 points (1 point on the occlusal, mesial triangular, and distal triangular fossae) on each file (Fig. 3). After superimposition, color maps that represented 3D deviations were generated with the maximum/minimum critical (nominal) values set at +50 μm and -50 μm and the tolerance range set at +10 μm and -10 μm.
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness and margin quality of 3D-printed ınterim dental crowns.
      ,
      • Yilmaz B.
      • Marques V.R.
      • Donmez M.B.
      • et al.
      Influence of 3D analysis software on measured deviations of CAD-CAM resin crowns from virtual design file: an in-vitro study.
      ,
      • Wang W.
      • Yu H.
      • Liu Y.
      • Jiang X.
      • Gao B.
      Trueness analysis of zirconia crowns fabricated with 3-dimensional printing.
      The deviation was analyzed by using the root mean square (RMS) method, the square root of the mean square of a set of numbers.
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness and margin quality of 3D-printed ınterim dental crowns.
      ,
      • Yilmaz B.
      • Marques V.R.
      • Donmez M.B.
      • et al.
      Influence of 3D analysis software on measured deviations of CAD-CAM resin crowns from virtual design file: an in-vitro study.
      STL files were imported again for the evaluation of other surfaces (external, occlusal, intaglio occlusal, and marginal). These surfaces were virtually separated, which divided the patterns into 4.
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness and margin quality of 3D-printed ınterim dental crowns.
      ,
      • Wang W.
      • Yu H.
      • Liu Y.
      • Jiang X.
      • Gao B.
      Trueness analysis of zirconia crowns fabricated with 3-dimensional printing.
      This superimposition process was repeated for each surface, and RMS values were automatically calculated from the color maps and given either as a positive (overcontoured areas) or a negative (undercontoured areas) value in the software program (Fig. 4). However, absolute values of the calculated RMS values were used for the statistical analyses.
      Figure thumbnail gr3
      Figure 3Points selected for superimposition process.
      Figure thumbnail gr4
      Figure 4Color maps generated by superimposing TC-STLs over C-STL for each surface. Red color indicates overcontoured and blue color indicates undercontoured areas considering nominal values set at +50 μm and -50 μm. Green color indicates acceptable areas considering tolerance range set at +10 μm and -10 μm. BC, Brilliant Crios; GR, G-CAM; MS, Crowntec; PMMA, breCAM.monoCOM.
      The normality of the data was analyzed by using the Kolmogorov-Smirnov test. Because of the nonnormal distribution, the Kruskal-Wallis and Dunn tests were performed by using a software program (IBM SPSS Statistics v22.0; IBM Corp) (α=.05).

      Results

      The Kruskal-Wallis tests revealed significant differences among the test groups for all surfaces analyzed (overall P<.001; external P<.001; occlusal P=.001; intaglio occlusal P<.001; marginal P<.001). Table 1 summarizes the descriptive statistics of each material-surface pair. Figure 5 illustrates the boxplot for RMS values (deviations). GR and PMMA crowns had similar overall (P=.441) and external (P=.541) deviations that were lower than those of MS crowns (P≤.007 for overall RMS and P≤.001 for external surface RMS). In addition, BC crowns had deviations similar to those of PMMA (P=.609) and MS (P=.178) when the external surface was concerned and to all groups when the overall RMS values were considered (P≥.052). However, BC crowns had higher external surface deviations than GR crowns (P=.005). For occlusal deviations, MS and GR crowns showed similar results (P>.05) that were lower than those of BC crowns (P≤.007). PMMA crowns had occlusal deviations similar to those of the other groups (P≥.317). While GR crowns had lower intaglio occlusal deviations than BC (P=.001) and PMMA (P=.028) crowns, MS crowns had lower deviations than BC crowns (P=.009). The differences between GR and MS (P>.05), PMMA and MS (P=.198), and PMMA and BC (P>.05) were not significant (P>.05). MS and GR crowns had higher marginal deviations than BC crowns (P≤.018), whereas PMMA crowns had deviations similar to those of the other groups (P≥.119).
      Table 1Descriptive statistics of RMS values (μm) according to each material-surface pair
      Material TypeSurface Type
      OverallExternalOcclusalIntaglio OcclusalMarginal
      Mean ±SDMedian (Min-Max)Mean ±SDMedian (Min-Max)Mean ±SDMedian (Min-Max)Mean ±SDMedian (Min-Max)Mean ±SDMedian (Min-Max)
      BC62 ±1663ab (42-86)58 ±1661bc (34-78)29 ±430b (23-35)31 ±933c (15-41)9 ±110a (7-11)
      PMMA83 ±11246a (36-400)39 ±1234ab (29-62)25 ±525ab (15-34)22 ±620bc (16-36)11 ±111ab (9-12)
      MS116 ±5115b (110-125)93 ±493c (87-98)20 ±621a (9-29)17 ±217ab (14-20)12 ±112b (11-12)
      GR38 ±635a (32-50)31 ±231a (28-33)20 ±520a (12-29)16 ±216a (13-19)14 ±214b (10-17)
      BC, Brilliant Crios; GR, G-CAM; MS, Crowntec; PMMA, breCAM.monoCOM.
      Different superscript letters indicate significate differences in columns (P<.05).
      Figure thumbnail gr5
      Figure 5Box-plot graph RMS values (μm) for each material-surface pair. BC, Brilliant Crios; GR, G-CAM; MS, Crowntec; PMMA, breCAM.monoCOM.

      Discussion

      Significant differences were observed among the test groups for overall deviations and for each surface analyzed. Therefore, the null hypothesis was rejected.
      Among the materials tested, MS crowns had significantly higher overall and external surface RMS values than GR and PMMA crowns. In addition, BC crowns had higher external surface RMS values than GR crowns. When color maps were further investigated (Fig. 4), MS differed noticeably from other materials as the blue color, which indicates undercontoured surfaces, was dominant. The undercontoured surfaces may be because of the difference in the manufacturing technique, as MS was the only additively manufactured material tested. Clinically, the undercontour can be interpreted as MS crowns having lighter interproximal contacts than other crowns, which might require a remake. The red color, which indicates overcontoured surfaces, was dominant on the occlusal surface of MS. The overcontouring found on the occlusal surfaces of MS may be because printing supports were positioned occlusally, and even though attention was paid when removing the supports, some support residue may have been left; this was not present with subtractively manufactured crowns. Considering that green was the primary color seen in GR and PMMA crowns and that BC crowns had predominantly blue on their occlusal surfaces, MS crowns may require more chairside time for occlusal adjustments. As for BC crowns, red was the predominant color on the external surface, which may lead to tighter interproximal contacts, as well as poor esthetics because of overcontouring. The higher external surface RMS values of BC crowns compared with those of GR crowns may be associated with their inherent material properties. The manufacturers of these materials have reported their elastic moduli as 10.3 GPa for BC and 3.2 GPa for GR., This difference might have led to the easier machinability of GR crowns. Nevertheless, considering that previous studies have reported better mechanical properties
      • Ciocan L.T.
      • Ghitman J.
      • Vasilescu V.G.
      • Iovu H.
      Mechanical properties of polymer-based blanks for machined dental restorations.
      ,
      • Agarwalla S.V.
      • Malhotra R.
      • Rosa V.
      Translucency, hardness and strength parameters of PMMA resin containing graphene-like material for CAD/CAM restorations.
      • Di Carlo S.
      • De Angelis F.
      • Brauner E.
      • et al.
      Flexural strength and elastic modulus evaluation of structures made by conventional PMMA and PMMA reinforced with graphene.
      • Hernández J.
      • Mora K.
      • Boquete-Castro A.
      • Kina S.
      The effect of thermocycling on surface microhardness of PMMA doped with graphene: an experimental in vitro study.
      and the fact that GR had deviations overall and on external surfaces similar to those of the PMMA tested in the present study, GR may provide longer clinical service. However, clinical studies are needed to substantiate this hypothesis.
      When occlusal and intaglio occlusal surface deviations were evaluated, BC crowns had significantly higher deviations than MS and GR crowns. In addition, PMMA crowns had intaglio occlusal surface deviations similar to those of other crowns, except for GR crowns. Color maps showed that, except for BC crowns, intaglio axial walls of the crowns were predominantly green. However, red was dominant at the axial walls of BC crowns; thus, intaglio adjustments may be required for BC crowns. Nevertheless, the maximum difference in mean deviation values at the occlusal surface was only 9 μm, and the maximum difference in mean deviation values at the intaglio occlusal surface was 15 μm among the materials. In addition, mean marginal RMS values ranged from 9 to 14 μm. Because these differences could be considered clinically small and because the mean deviations of all materials at the occlusal and intaglio occlusal surfaces were generally smaller than the cement space of 30 μm, the clinical fit of the crowns fabricated by using these materials may be similar. However, this interpretation needs support from clinical trials. The intaglio surface trueness of interim crowns fabricated either with additive or subtractive manufacturing has been investigated previously,
      • Son K.
      • Lee J.H.
      • Lee K.B.
      Comparison of intaglio surface trueness of interim dental crowns fabricated with SLA 3D printing, DLP 3D printing, and milling technologies.
      and subtractively manufactured PMMA crowns had significantly higher deviations for all surfaces tested (overall intaglio, marginal, axial, and occlusal), an outcome that contradicts the results of the present study. Even though both studies used PMMA for subtractive manufacturing, the additively manufactured resin tested in the present study is marketed as a definitive and has a different chemical structure compared with the interim additively manufactured materials.
      An IOS, which was reported to have a performance similar to that of a laboratory scanner,
      • Mangano F.
      • Lerner H.
      • Margiani B.
      • Solop I.
      • Latuta N.
      • Admakin O.
      Congruence between meshes and library files of ımplant scanbodies: an ın vitro study comparing five ıntraoral scanners.
      was used in the present study to digitize the crowns, as these scanners enable data acquisition of both intaglio and outer surfaces in 1 continuous motion. Even though laboratory scanners have been reported to have a higher accuracy than IOSs,
      • Pan Y.
      • Tsoi J.K.H.
      • Lam W.Y.H.
      • Pow E.H.N.
      Implant framework misfit: a systematic review on assessment methods and clinical complications.
      these scanners digitally stitch separate scans of intaglio and outer surfaces with their algorithm, which may have led to amplified deviations.
      • Çakmak G.
      • Donmez M.B.
      • Cuellar A.R.
      • Kahveci Ç.
      • Schimmel M.
      • Yilmaz B.
      Additive or subtractive manufacturing of crown patterns used for pressing or casting: a trueness analysis.
      Similar to previous studies on the trueness of additively manufactured prostheses,
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness of 3-unit interim fixed partial dentures.
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness and margin quality of 3D-printed ınterim dental crowns.
      • Çakmak G.
      • Donmez M.B.
      • Cuellar A.R.
      • Kahveci Ç.
      • Schimmel M.
      • Yilmaz B.
      Additive or subtractive manufacturing of crown patterns used for pressing or casting: a trueness analysis.
      3-point alignment was used for the superimposition of STL files, while a freeware and the RMS method were used for the trueness analysis. However, different alignment algorithms,
      • O'Toole S.
      • Osnes C.
      • Bartlett D.
      • Keeling A.
      Investigation into the accuracy and measurement methods of sequential 3D dental scan alignment.
      3D analysis software programs,
      • Cakmak G.
      • Marques V.R.
      • Donmez M.B.
      • Lu W.E.
      • Abou-Ayash S.
      • Yilmaz B.
      Comparison of measured deviations in digital implant scans depending on software and operator.
      or deviation measurement methods
      • Lerner H.
      • Nagy K.
      • Pranno N.
      • Zarone F.
      • Admakin O.
      • Mangano F.
      Trueness and precision of 3D-printed versus milled monolithic zirconia crowns: an in vitro study.
      may lead to different results.
      Even though significant differences were observed and the number of specimens in each group was based on the results of previous studies,
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness of 3-unit interim fixed partial dentures.
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness and margin quality of 3D-printed ınterim dental crowns.
      • Çakmak G.
      • Donmez M.B.
      • Cuellar A.R.
      • Kahveci Ç.
      • Schimmel M.
      • Yilmaz B.
      Additive or subtractive manufacturing of crown patterns used for pressing or casting: a trueness analysis.
      ,
      • Lerner H.
      • Nagy K.
      • Pranno N.
      • Zarone F.
      • Admakin O.
      • Mangano F.
      Trueness and precision of 3D-printed versus milled monolithic zirconia crowns: an in vitro study.
      the absence of a power analysis is a limitation. A single operator performed test scans in the same room and under the same conditions for standardization; different operators
      • Cakmak G.
      • Marques V.R.
      • Donmez M.B.
      • Lu W.E.
      • Abou-Ayash S.
      • Yilmaz B.
      Comparison of measured deviations in digital implant scans depending on software and operator.
      and ambient conditions
      • Revilla-León M.
      • Jiang P.
      • Sadeghpour M.
      • et al.
      Intraoral digital scans-part 1: influence of ambient scanning light conditions on the accuracy (trueness and precision) of different intraoral scanners.
      ,
      • Revilla-León M.
      • Jiang P.
      • Sadeghpour M.
      • et al.
      Intraoral digital scans: part 2-influence of ambient scanning light conditions on the mesh quality of different intraoral scanners.
      may affect the results. Only one 3D printer and resin were used; however, previous studies have reported significant differences between different additively manufacturing techniques
      • Son K.
      • Lee J.H.
      • Lee K.B.
      Comparison of intaglio surface trueness of interim dental crowns fabricated with SLA 3D printing, DLP 3D printing, and milling technologies.
      and various other parameters such as build angle, layer thickness, laser intensity, and laser speed; the geometry of the supporting structures may also affect the efficiency of additive manufacturing.
      • Çakmak G.
      • Cuellar A.R.
      • Donmez M.B.
      • et al.
      Effect of printing layer thickness on the trueness and margin quality of 3D-printed ınterim dental crowns.
      In addition, 1 type of prosthesis was investigated, and an increased number of retainers or pontics may lead to higher deviations.
      • Moon W.
      • Kim S.
      • Lim B.S.
      • Park Y.S.
      • Kim R.J.
      • Chung S.H.
      Dimensional accuracy evaluation of temporary dental restorations with different 3D printing systems.
      The trueness of a prosthesis is only one of the parameters that affect clinical longevity. Therefore, in vitro and in vivo investigations of the other mechanical and optical properties of GR and MS, such as their wear resistance, fracture resistance, color stability, and translucency, would broaden the knowledge on the applicability of these materials in clinical situations given that they are relatively new when compared with the other tested materials.

      Conclusions

      Based on the findings of this in vitro study, the following conclusions were drawn:
      • 1.
        Material type had a significant effect on the trueness of crowns.
      • 2.
        The tested additively manufactured composite resin crowns had overall and external surface deviations that were either similar to or higher than those of subtractively manufactured crowns. Combined with the information obtained through color maps, these deviations may indicate more chairside occlusal adjustments and lighter interproximal contacts for tested additively manufactured composite resin crowns, which might even lead to a remake in clinical situations.
      • 3.
        The clinical fit of the crowns fabricated from the tested materials may be similar, as deviations from the virtual design file at the occlusal, intaglio occlusal, and marginal surfaces can be considered small.

      CRediT authorship contribution statement

      Gülce Çakmak: Conceptualization, Methodology, Investigation. Ana Maria Rusa: Methodology, Investigation. Mustafa Borga Donmez: Writing – original draft. Canan Akay: Resources, Formal analysis. Martin Schimmel: Conceptualization, Methodology, Supervision. Burak Yilmaz: Conceptualization, Methodology, Supervision, Writing – review & editing.

      Acknowledgments

      The authors thank Saremco Dental AG and Coltène AG for supplying the materials used in this study. Ivoclar AG is also gratefully acknowledged for milling support. The authors also thank Marco Pannier for his contributions to the specimen preparation. The authors dedicate this article to the contributing memory of Dr. Ralf Böhner, Coltène AG, who passed away on October 2021.

      References

        • Revilla-León M.
        • Jiang P.
        • Sadeghpour M.
        • et al.
        Intraoral digital scans-part 1: influence of ambient scanning light conditions on the accuracy (trueness and precision) of different intraoral scanners.
        J Prosthet Dent. 2020; 124: 372-378
        • Revilla-León M.
        • Jiang P.
        • Sadeghpour M.
        • et al.
        Intraoral digital scans: part 2-influence of ambient scanning light conditions on the mesh quality of different intraoral scanners.
        J Prosthet Dent. 2020; 124: 575-580
        • Cakmak G.
        • Marques V.R.
        • Donmez M.B.
        • Lu W.E.
        • Abou-Ayash S.
        • Yilmaz B.
        Comparison of measured deviations in digital implant scans depending on software and operator.
        J Dent. 2022; 122: 104154
        • Beuer F.
        • Schweiger J.
        • Edelhoff D.
        Digital dentistry: an overview of recent developments for CAD/CAM generated restorations.
        Br Dent J. 2008; 204: 505-511
        • Son K.
        • Lee J.H.
        • Lee K.B.
        Comparison of intaglio surface trueness of interim dental crowns fabricated with SLA 3D printing, DLP 3D printing, and milling technologies.
        Healthcare (Basel). 2021; 9: 983
        • Moon W.
        • Kim S.
        • Lim B.S.
        • Park Y.S.
        • Kim R.J.
        • Chung S.H.
        Dimensional accuracy evaluation of temporary dental restorations with different 3D printing systems.
        Materials (Basel). 2021; 14: 1487
        • Alharbi N.
        • Osman R.B.
        • Wismeijer D.
        Factors ınfluencing the dimensional accuracy of 3D-printed full-coverage dental restorations using stereolithography technology.
        Int J Prosthodont. 2016; 29: 503-510
        • Çakmak G.
        • Cuellar A.R.
        • Donmez M.B.
        • et al.
        Effect of printing layer thickness on the trueness of 3-unit interim fixed partial dentures.
        J Prosthet Dent. 27 May 2022; ([Epub ahead of print])https://doi.org/10.1016/j.prosdent.2022.04.015
        • Çakmak G.
        • Cuellar A.R.
        • Donmez M.B.
        • et al.
        Effect of printing layer thickness on the trueness and margin quality of 3D-printed ınterim dental crowns.
        Appl Sci. 2021; 11: 9246
        • Çakmak G.
        • Donmez M.B.
        • Cuellar A.R.
        • Kahveci Ç.
        • Schimmel M.
        • Yilmaz B.
        Additive or subtractive manufacturing of crown patterns used for pressing or casting: a trueness analysis.
        J Dent. 2022; 124: 104221
        • Shim J.S.
        • Kim J.E.
        • Jeong S.H.
        • Choi Y.J.
        • Ryu J.J.
        Printing accuracy, mechanical properties, surface characteristics, and microbial adhesion of 3D-printed resins with various printing orientations.
        J Prosthet Dent. 2020; 124: 468-475
        • Yu B.Y.
        • Son K.
        • Lee K.B.
        Evaluation of intaglio surface trueness and margin quality of interim crowns in accordance with the build angle of stereolithography apparatus 3-dimensional printing.
        J Prosthet Dent. 2021; 126: 231-237
        • Lerner H.
        • Nagy K.
        • Pranno N.
        • Zarone F.
        • Admakin O.
        • Mangano F.
        Trueness and precision of 3D-printed versus milled monolithic zirconia crowns: an in vitro study.
        J Dent. 2021; 113: 103792
        • Peng C.C.
        • Chung K.H.
        • Yau H.T.
        • Ramos Jr., V.
        Assessment of the internal fit and marginal integrity of interim crowns made by different manufacturing methods.
        J Prosthet Dent. 2020; 123: 514-522
        • Guo N.
        • Leu M.C.
        Additive manufacturing: technology, applications and research needs.
        Front Mech Eng. 2013; 8: 215-243
        • Ciocan L.T.
        • Ghitman J.
        • Vasilescu V.G.
        • Iovu H.
        Mechanical properties of polymer-based blanks for machined dental restorations.
        Materials (Basel). 2021; 14: 7293
        • Corbani K.
        • Hardan L.
        • Eid R.
        • et al.
        Fracture resistance of three-unit fixed dental prostheses fabricated with milled and 3D printed composite-based materials.
        J Contemp Dent Pract. 2021; 22: 985-990
        • Agarwalla S.V.
        • Malhotra R.
        • Rosa V.
        Translucency, hardness and strength parameters of PMMA resin containing graphene-like material for CAD/CAM restorations.
        J Mech Behav Biomed Mater. 2019; 100: 103388
        • Di Carlo S.
        • De Angelis F.
        • Brauner E.
        • et al.
        Flexural strength and elastic modulus evaluation of structures made by conventional PMMA and PMMA reinforced with graphene.
        Eur Rev Med Pharmacol Sci. 2020; 24: 5201-5208
        • Hernández J.
        • Mora K.
        • Boquete-Castro A.
        • Kina S.
        The effect of thermocycling on surface microhardness of PMMA doped with graphene: an experimental in vitro study.
        J Clin Dent Res. 2020; 17: 152-161
        • Lee J.-H.
        • Jo J.-K.
        • Kim D.-A.
        • Patel K.D.
        • Kim H.-W.
        • Lee H.-H.
        Nano-graphene oxide incorporated into PMMA resin to prevent microbial adhesion.
        Dent Mater. 2018; 34: e63-e72
      1. The graphenano dental.
        (Available at:)
        • Grzebieluch W.
        • Kowalewski P.
        • Grygier D.
        • Rutkowska-Gorczyca M.
        • Kozakiewicz M.
        • Jurczyszyn K.
        Printable and machinable dental restorative composites for CAD/CAM application—comparison of mechanical properties, fractographic, texture and fractal dimension analysis.
        Materials. 2021; 14: 4919
        • Corbani K.
        • Hardan L.
        • Skienhe H.
        • Özcan M.
        • Alharbi N.
        • Salameh Z.
        Effect of material thickness on the fracture resistance and failure pattern of 3D-printed composite crowns.
        Int J Comput Dent. 2020; 23: 225-233
        • Zimmermann M.
        • Ender A.
        • Attin T.
        • Mehl A.
        Fracture load of three-unit full-contour fixed dental prostheses fabricated with subtractive and additive CAD/CAM technology.
        Clin Oral Investig. 2020; 24: 1035-1042
        • Zimmermann M.
        • Ender A.
        • Egli G.
        • Özcan M.
        • Mehl A.
        Fracture load of CAD/CAM-fabricated and 3D-printed composite crowns as a function of material thickness.
        Clin Oral Investig. 2019; 23: 2777-2784
        • Atria P.J.
        • Bordin D.
        • Marti F.
        • et al.
        3D-printed resins for provisional dental restorations: comparison of mechanical and biological properties.
        J Esthet Restor Dent. 2022; 34: 804-815
        • Donmez M.B.
        • Okutan Y.
        Marginal gap and fracture resistance of implant-supported 3D-printed definitive composite crowns: an in vitro study.
        J Dent. 2022; 124: 104216
        • Pacquet W.
        • Tapie L.
        • Mawussi B.
        • Boitelle P.
        Volumetric and dimensional accuracy assessment of CAD-CAM–manufactured dental prostheses from different materials.
        J Prosthet Dent. 27 July 2021; ([Epub ahead of print])https://doi.org/10.1016/j.prosdent.2021.05.024
        • Mangano F.
        • Lerner H.
        • Margiani B.
        • Solop I.
        • Latuta N.
        • Admakin O.
        Congruence between meshes and library files of ımplant scanbodies: an ın vitro study comparing five ıntraoral scanners.
        J Clin Med. 2020; 9: 2174
        • Yilmaz B.
        • Marques V.R.
        • Donmez M.B.
        • et al.
        Influence of 3D analysis software on measured deviations of CAD-CAM resin crowns from virtual design file: an in-vitro study.
        J Dent. 2022; 118: 103933
        • Wang W.
        • Yu H.
        • Liu Y.
        • Jiang X.
        • Gao B.
        Trueness analysis of zirconia crowns fabricated with 3-dimensional printing.
        J Prosthet Dent. 2019; 121: 285-291
      2. The coltène.
        (Available at:)
      3. The graphenano dental.
        (Available at:)
        • Pan Y.
        • Tsoi J.K.H.
        • Lam W.Y.H.
        • Pow E.H.N.
        Implant framework misfit: a systematic review on assessment methods and clinical complications.
        Clin Implant Dent Relat Res. 2021; 23: 244-258
        • O'Toole S.
        • Osnes C.
        • Bartlett D.
        • Keeling A.
        Investigation into the accuracy and measurement methods of sequential 3D dental scan alignment.
        Dent Mater. 2019; 35: 495-500