Research and Education| Volume 129, ISSUE 3, P478-485, March 2023

Effect of maximum support attachment angle on intaglio surface trueness of anatomic contour monolithic prostheses manufactured by digital light processing and zirconia suspension

Open AccessPublished:January 06, 2023DOI:https://doi.org/10.1016/j.prosdent.2022.12.003

      Abstract

      Statement of problem

      Support structures are essential for the quality of resin-based prostheses made by the digital light processing (DLP), but few studies have evaluated the effect of support structure on the accuracy of zirconia-based anatomic contour prostheses.

      Purpose

      The purpose of this in vitro study was to evaluate the effect of maximum support attachment angle (MSA) on the intaglio surface trueness of anatomic contour prostheses made by DLP and compare the trueness of 2-unit anatomic contour prostheses with that of those produced by milling.

      Material and methods

      Anatomic contour single-unit prostheses were manufactured using DLP and a suspension with 3-mol% yttria-stabilized zirconia. Four different conditions of MSA values to the vertical axis of the object (50, 55, 60, and 65 degrees) were applied (n=10). After printing, postprocessing, and sintering, all successfully produced prostheses were evaluated for intaglio surface trueness by considering the root mean square (RMS). Using the MSA showing the highest trueness, the 2-unit prostheses made by DLP (DLP group) were compared with milled (MIL group) prostheses in terms of intaglio accuracy (n=10). One-way analysis of variance and a post hoc pairwise comparison or independent t test were used for trueness analysis (α=.05).

      Results

      Three MSA groups (50, 55, and 60 degrees) were successfully produced with significant differences between the trueness of the single-unit prostheses for the groups with different MSA values (P<.05). The highest trueness was in the 50-degree MSA group. The 2-unit prostheses of the DLP group with 50-degree MSA showed significantly lower trueness than those of the MIL group (P<.05); however, the RMS values of both groups were lower than 50 μm.

      Conclusions

      The intaglio surface trueness of anatomic contour DLP-generated prostheses can be improved by changing the MSA. The 50-degree MSA was beneficial for the accuracy of both single-unit and 2-unit DLP-generated prostheses, produced within clinically acceptable limits.
      Clinical Implications
      Considering the MSA, additional support structures positioned over the occlusal areas of anatomic contour prostheses can improve the intaglio surface accuracy of DLP-generated zirconia single-unit and 2-unit zirconia prostheses to a clinically acceptable range.
      With advances in computer-aided design and computer-aided manufacturing (CAD-CAM) technology, subtractive manufacturing has been considered the standard for fabricating ceramic prostheses. However, limited design freedom, including thin-walled features or small radii, high consumption of raw materials, considerable waste from unused remnants of milled blocks, and severe wear of tools are some of the concerns of subtractive manufacturing.
      • Baumgartner S.
      • Gmeiner R.
      • Schönherr J.A.
      • Stampfl J.
      Stereolithography-based additive manufacturing of lithium disilicate glass ceramic for dental applications.
      ,
      • Lebon N.
      • Tapie L.
      • Duret F.
      • Attal J.-P.
      Understanding dental CAD/CAM for restorations-dental milling machines from a mechanical engineering viewpoint. Part A: chairside milling machines.
      An anatomic contour monolithic prosthesis has a thin-walled ceramic shell structure with complex geometric shapes, including grooves, cusps, and fossae. Machining complex surfaces with high precision and reproducibility is challenging because of the limited access to smaller areas imposed by the tool size and machining axis.
      • Lebon N.
      • Tapie L.
      • Duret F.
      • Attal J.-P.
      Understanding dental CAD/CAM for restorations-dental milling machines from a mechanical engineering viewpoint. Part A: chairside milling machines.
      ,
      • Lian Q.
      • Wu X.
      • Li D.
      • et al.
      Accurate printing of a zirconia molar crown bridge using three-part auxiliary supports and ceramic mask projection stereolithography.
      In contrast, additive manufacturing (AM) is advantageous in terms of having almost no sensitivity to complex-shaped structures, no tool wear, and custom production.
      • Lian Q.
      • Wu X.
      • Li D.
      • et al.
      Accurate printing of a zirconia molar crown bridge using three-part auxiliary supports and ceramic mask projection stereolithography.
      Researchers have recently focused on ceramic AM to construct complex structures such as anatomic contour monolithic prostheses by using properly formulated materials and corresponding printers.
      • Baumgartner S.
      • Gmeiner R.
      • Schönherr J.A.
      • Stampfl J.
      Stereolithography-based additive manufacturing of lithium disilicate glass ceramic for dental applications.
      ,
      • Lian Q.
      • Wu X.
      • Li D.
      • et al.
      Accurate printing of a zirconia molar crown bridge using three-part auxiliary supports and ceramic mask projection stereolithography.
      • Li H.
      • Song L.
      • Sun J.
      • Ma J.
      • Shen Z.
      Dental ceramic prostheses by stereolithography-based additive manufacturing: potentials and challenges.
      • Schönherr J.A.
      • Baumgartner S.
      • Hartmann M.
      • Stampfl J.
      Stereolithographic additive manufacturing of high precision glass ceramic parts.
      • Revilla-León M.
      • Meyer M.J.
      • Zandinejad A.
      • Özcan M.
      Additive manufacturing technologies for processing zirconia in dental applications.
      • Li R.
      • Xu T.
      • Wang Y.
      • Sun Y.
      Accuracy of zirconia crowns manufactured by stereolithography with an occlusal full-supporting structure: An in vitro study.
      • Osman R.B.
      • van der Veen A.J.
      • Huiberts D.
      • Wismeijer D.
      • Alharbi N.
      3D-printing zirconia implants; a dream or a reality? An in-vitro study evaluating the dimensional accuracy, surface topography and mechanical properties of printed zirconia implant and discs.
      • Li R.
      • Wang Y.
      • Hu M.
      • et al.
      Strength and adaptation of stereolithography-fabricated zirconia dental crowns: an in vitro study.
      • Revilla-León M.
      • Methani M.M.
      • Morton D.
      • Zandinejad A.
      Internal and marginal discrepancies associated with stereolithography (SLA) additively manufactured zirconia crowns.
      • Wang W.
      • Sun J.
      Dimensional accuracy and clinical adaptation of ceramic crowns fabricated with the stereolithography technique.
      • Wang W.
      • Yu H.
      • Liu Y.
      • Jiang X.
      • Gao B.
      Trueness analysis of zirconia crowns fabricated with 3-dimensional printing.
      A photosensitive suspension with an adequate solid loading of zirconia and homogeneous dispersion is fundamental.
      • Griffith M.L.
      • Halloran J.W.
      Freeform fabrication of ceramics via stereolithography.
      At present, 3-mol% yttria-stabilized tetragonal zirconia polycrystal (3Y-TZP) is extensively used for dental restorations and is the material of choice for AM.
      • Li H.
      • Song L.
      • Sun J.
      • Ma J.
      • Shen Z.
      Dental ceramic prostheses by stereolithography-based additive manufacturing: potentials and challenges.
      ,
      • Khanlar L.N.
      • Salazar Rios A.
      • Tahmaseb A.
      • Zandinejad A.
      Additive manufacturing of zirconia ceramic and its application in clinical dentistry: a review.
      Among the various technical strategies for AM with zirconia, vat photopolymerization requires a mixture of liquid resin and ceramic powders to manufacture a complex object by selectively solidifying the ceramic suspension through controlled light-induced polymerization.
      • Revilla-León M.
      • Meyer M.J.
      • Zandinejad A.
      • Özcan M.
      Additive manufacturing technologies for processing zirconia in dental applications.
      During vat polymerization, printing parameters, support parameters, and postprocessing protocols have been reported to influence the properties of AM objects.
      • Piedra-Cascón W.
      • Krishnamurthy V.R.
      • Att W.
      • Revilla-León M.
      3D printing parameters, supporting structures, slicing, and post-processing procedures of vat-polymerization additive manufacturing technologies: a narrative review.
      Studies using stereolithography (SLA) or digital light processing (DLP) techniques have reported the potential and challenges of fabricating zirconia-based dental prostheses, mainly focusing on strength, density, and trueness.
      • Lian Q.
      • Wu X.
      • Li D.
      • et al.
      Accurate printing of a zirconia molar crown bridge using three-part auxiliary supports and ceramic mask projection stereolithography.
      • Li H.
      • Song L.
      • Sun J.
      • Ma J.
      • Shen Z.
      Dental ceramic prostheses by stereolithography-based additive manufacturing: potentials and challenges.
      • Schönherr J.A.
      • Baumgartner S.
      • Hartmann M.
      • Stampfl J.
      Stereolithographic additive manufacturing of high precision glass ceramic parts.
      • Revilla-León M.
      • Meyer M.J.
      • Zandinejad A.
      • Özcan M.
      Additive manufacturing technologies for processing zirconia in dental applications.
      • Li R.
      • Xu T.
      • Wang Y.
      • Sun Y.
      Accuracy of zirconia crowns manufactured by stereolithography with an occlusal full-supporting structure: An in vitro study.
      • Osman R.B.
      • van der Veen A.J.
      • Huiberts D.
      • Wismeijer D.
      • Alharbi N.
      3D-printing zirconia implants; a dream or a reality? An in-vitro study evaluating the dimensional accuracy, surface topography and mechanical properties of printed zirconia implant and discs.
      • Li R.
      • Wang Y.
      • Hu M.
      • et al.
      Strength and adaptation of stereolithography-fabricated zirconia dental crowns: an in vitro study.
      • Revilla-León M.
      • Methani M.M.
      • Morton D.
      • Zandinejad A.
      Internal and marginal discrepancies associated with stereolithography (SLA) additively manufactured zirconia crowns.
      • Wang W.
      • Sun J.
      Dimensional accuracy and clinical adaptation of ceramic crowns fabricated with the stereolithography technique.
      • Wang W.
      • Yu H.
      • Liu Y.
      • Jiang X.
      • Gao B.
      Trueness analysis of zirconia crowns fabricated with 3-dimensional printing.
      DLP has higher manufacturing speed and resolution than SLA because of fast shifting and integral projecting.
      • Lee M.P.
      • Cooper G.J.
      • Hinkley T.
      • Gibson G.M.
      • Padgett M.J.
      • Cronin L.
      Development of a 3D printer using scanning projection stereolithography.
      Complex ceramic structures with a high degree of detail can be produced using DLP.
      • Khanlar L.N.
      • Salazar Rios A.
      • Tahmaseb A.
      • Zandinejad A.
      Additive manufacturing of zirconia ceramic and its application in clinical dentistry: a review.
      However, the quality of AM ceramic objects is dependent on the suspension composition, printing parameters, and support structures.
      • Li H.
      • Song L.
      • Sun J.
      • Ma J.
      • Shen Z.
      Dental ceramic prostheses by stereolithography-based additive manufacturing: potentials and challenges.
      • Schönherr J.A.
      • Baumgartner S.
      • Hartmann M.
      • Stampfl J.
      Stereolithographic additive manufacturing of high precision glass ceramic parts.
      • Revilla-León M.
      • Meyer M.J.
      • Zandinejad A.
      • Özcan M.
      Additive manufacturing technologies for processing zirconia in dental applications.
      • Li R.
      • Xu T.
      • Wang Y.
      • Sun Y.
      Accuracy of zirconia crowns manufactured by stereolithography with an occlusal full-supporting structure: An in vitro study.
      ,
      • Silva N.R.
      • Witek L.
      • Coelho P.G.
      • Thompson V.P.
      • Rekow E.D.
      • Smay J.
      Additive CAD/CAM process for dental prostheses.
      Support structures are important in terms of both the accuracy and position of AM objects.
      • Li H.
      • Song L.
      • Sun J.
      • Ma J.
      • Shen Z.
      Dental ceramic prostheses by stereolithography-based additive manufacturing: potentials and challenges.
      • Schönherr J.A.
      • Baumgartner S.
      • Hartmann M.
      • Stampfl J.
      Stereolithographic additive manufacturing of high precision glass ceramic parts.
      • Revilla-León M.
      • Meyer M.J.
      • Zandinejad A.
      • Özcan M.
      Additive manufacturing technologies for processing zirconia in dental applications.
      • Li R.
      • Xu T.
      • Wang Y.
      • Sun Y.
      Accuracy of zirconia crowns manufactured by stereolithography with an occlusal full-supporting structure: An in vitro study.
      The support structure and build orientation significantly affect the dimensional accuracy of AM polymer objects.
      • Piedra-Cascón W.
      • Krishnamurthy V.R.
      • Att W.
      • Revilla-León M.
      3D printing parameters, supporting structures, slicing, and post-processing procedures of vat-polymerization additive manufacturing technologies: a narrative review.
      ,
      • Alharbi N.
      • Osman R.B.
      • Wismeijer D.
      Factors influencing the dimensional accuracy of 3D-printed full-coverage dental restorations using stereolithography technology.
      During ceramic AM, the support structure attached to the printing object enables the construction of layers on the build platform, mainly for overhanging areas.
      • Schönherr J.A.
      • Baumgartner S.
      • Hartmann M.
      • Stampfl J.
      Stereolithographic additive manufacturing of high precision glass ceramic parts.
      To fully reproduce the intaglio and occlusal surfaces of dental prostheses, the optimal use of support structures for ceramic AM is essential.
      • Lian Q.
      • Wu X.
      • Li D.
      • et al.
      Accurate printing of a zirconia molar crown bridge using three-part auxiliary supports and ceramic mask projection stereolithography.
      ,
      • Silva N.R.
      • Witek L.
      • Coelho P.G.
      • Thompson V.P.
      • Rekow E.D.
      • Smay J.
      Additive CAD/CAM process for dental prostheses.
      The geometric design of supports allows complex structures to be reproduced and minimizes residues after postprocessing.
      • Lian Q.
      • Wu X.
      • Li D.
      • et al.
      Accurate printing of a zirconia molar crown bridge using three-part auxiliary supports and ceramic mask projection stereolithography.
      The accuracy of the prosthesis in the z-direction has been reported to be improved by adding supports.
      • Lian Q.
      • Wu X.
      • Li D.
      • et al.
      Accurate printing of a zirconia molar crown bridge using three-part auxiliary supports and ceramic mask projection stereolithography.
      Trueness can be defined as the performance of the manufacturing process in reproducing an object as closely as possible to its virtual form.
      • Braian M.
      • Jimbo R.
      • Wennerberg A.
      Production tolerance of additive manufactured polymeric objects for clinical applications.
      The intaglio surface trueness of the ceramic AM crown has been reported to be affected by the support structure, showing low precision or reproducibility in the case of inadequate shapes or dimensions.
      • Schönherr J.A.
      • Baumgartner S.
      • Hartmann M.
      • Stampfl J.
      Stereolithographic additive manufacturing of high precision glass ceramic parts.
      ,
      • Li R.
      • Xu T.
      • Wang Y.
      • Sun Y.
      Accuracy of zirconia crowns manufactured by stereolithography with an occlusal full-supporting structure: An in vitro study.
      ,
      • Silva N.R.
      • Witek L.
      • Coelho P.G.
      • Thompson V.P.
      • Rekow E.D.
      • Smay J.
      Additive CAD/CAM process for dental prostheses.
      Therefore, the purpose of this in vitro study was to evaluate the effect of the maximum angle of the support attachment to the vertical axis on the intaglio surface trueness of anatomic contour monolithic prostheses made by using DLP and a zirconia suspension. The research hypotheses were that the changes in maximum angle of support attachment would affect the intaglio surface trueness of the single-unit zirconia-based prosthesis made by using DLP and that the intaglio surface trueness of the 2-unit anatomic contour dental prosthesis would be different between the milling and the DLP after applying the optimal value of the maximum angle of support attachment.

      Material and methods

      An acrylic resin mandibular right first molar (Simple Root Tooth Model; Nissin Dental Products) was prepared for an anatomic contour restoration with 1.5-mm occlusal reduction, 1 to 1.5-mm axial reduction with rounded internal line angles, and a 1-mm circumferential chamfer finish line. The prepared tooth was digitized with a laboratory scanner (T500; Medit). A virtual anatomic contour single-unit prosthesis was designed (reference CAD #1) (Dental Designer; 3Shape A/S) with a cement space of 25 μm starting from 1 mm to the finish line. Acrylic resin maxillary right first and second molars (Simple Root Tooth Model; Nissin Dental Products) were also prepared using dimensions similar to those for anatomic contour restorations and scanned. An anatomic contour 2-unit prosthesis was virtually designed (reference CAD #2) on the prepared abutments with the same cement space as the single-unit prosthesis. These two reference CAD data were used to produce the specimens in the study.
      For the DLP process, the support structures were attached to the thick occlusal surface of each prosthesis by using a slicing software program (MetaMorp DLP Slicer; MetaMorp).
      • Lian Q.
      • Wu X.
      • Li D.
      • et al.
      Accurate printing of a zirconia molar crown bridge using three-part auxiliary supports and ceramic mask projection stereolithography.
      The grooves, fossae, and cusp tips were excluded from the region of support, minimizing interference with the occlusal contour of the prosthesis. The intaglio surface remained intact without the support structure to ensure fit to the abutment. Following the geometric shapes of the occlusal surface, the maximum support attachment angle (MSA) was defined as the maximum angle (degree) formed between the vertical axis of the prosthesis and the area of attached supports viewed in the vertical plane and was evaluated in this study (Fig. 1). Considering the overhang area to be higher than a 45-degree angle to the vertical axis, 4 different experimental MSA values were tested to determine the highest trueness of the anatomic contour single-unit prosthesis (reference CAD #1) as follows: 50, 55, 60, and 65 degrees. Each MSA implied that the area of support attachment was distributed up to the value. In other words, the smaller the MSA, the wider the area of support attachment spread on the surface to be printed. An MSA greater than 70 degrees was excluded from the experiment so that the support structures would not be attached near the cusp tip areas. The structural unit of each support was attached perpendicular to the surface of the prosthesis, with a maximum diameter of 0.3 mm. The distance between each support structure was uniformly set as 1.0 mm. To exclude the possible effect of build orientation, each prosthesis was layered at a 0-degree angulation, with its long axis perpendicular to the build platform. Trueness analysis was performed with a clinically acceptable limit, 50 μm for the single-unit prosthesis in an absolute evaluation manner and without a control group as in previous studies.
      • Wang W.
      • Sun J.
      Dimensional accuracy and clinical adaptation of ceramic crowns fabricated with the stereolithography technique.
      ,
      • Peters M.C.
      • Delong R.
      • Pintado M.R.
      • Pallesen U.
      • Qvist V.
      • Douglas W.H.
      Comparison of two measurement techniques for clinical wear.
      • Cho J.H.
      • Yoon H.I.
      • Han J.S.
      • Kim D.J.
      Trueness of the inner surface of monolithic crowns fabricated by milling of a fully sintered (Y, Nb)-TZP block in Chairside CAD-CAM system for single-visit dentistry.
      • Rungrojwittayakul O.
      • Kan J.Y.
      • Shiozaki K.
      • et al.
      Accuracy of 3D Printed Models Created by Two Technologies of Printers with Different Designs of Model Base.
      • Al Hamad K.Q.
      • Al-Rashdan B.A.
      • Ayyad J.Q.
      • et al.
      Additive manufacturing of dental ceramics: a systematic review and meta-analysis.
      • Kim Y.K.
      • Han J.S.
      • Yoon H.I.
      Evaluation of intaglio surface trueness, wear, and fracture resistance of zirconia crown under simulated mastication: a comparative analysis between subtractive and additive manufacturing.
      Figure thumbnail gr1
      Figure 1Schematic two-dimensional diagram of support structures attached perpendicular to surface of occlusal area of anatomic contour crown, viewed in vertical plane. Considering overhang area, 4 different MSA values (50, 55, 60, and 65 degrees) greater than 45 degrees from vertical axis of prosthesis evaluated. MSA, maximum support attachment angle.
      For each design group of support structures, 10 anatomic contour single-unit zirconia-based prostheses were fabricated by DLP using the reference CAD #1. A zirconia suspension (Cera-P; M.O.P) containing nanometer-sized 3Y-TZP powder, polyfunctional acrylates, and dispersant was used (Table 1). The scaling factor was calculated from the estimated shrinkage rate of the AM object after sintering, 25%±1% in every direction. A DLP printer with a 3-μm resolution laser measurement sensor (Octave Light R1; Octave Light Ltd) was used to produce the prostheses with a top-down approach. The printing parameters were 50-μm layer thickness, 40×40-μm X-Y plane resolution, 5-μm Z-axis resolution, 405-nm wavelength of ultraviolet light-emitting diodes, 150-mJ exposure intensity, exposure time of 5 seconds, and 30% decrease in blade speed from the default printer setting. For each design group, the prostheses were manufactured simultaneously by using the same batch of processes located in the center of the build platform. After printing, all support structures were manually removed from the surfaces of the AM objects. The procedures were carefully performed to prevent interference with the AM objects. The AM objects were cleaned with ethanol, dried, and treated for 6 hours to a temperature of 400 °C for debinding, according to the manufacturer’s instructions. Subsequently, the AM object was sintered to full density for several hours, with an increasing rate of 10 °C/min and a 2-hour holding time at 1450 °C. The intaglio surfaces of the prostheses remained intact without any adjustments before and after postprocessing. If the prosthesis had any structural defect or surface damage after postprocessing, or if there were any types of directly detectable structural failures during printing (delamination or detachment), they were excluded from further analysis.
      Table 1Composition of zirconia suspension used for digital light processing
      ComponentWeight%
      Hexanediol diacrylate5.0
      Isobornyl acrylate1.0
      Trifunctional acrylate3.8
      Dispersing agent5.0
      Photoinitiator0.2
      Di(propylene glycol)methyl ether5.0
      3 mol% yttria-stabilized tetragonal zirconia polycrystal80.0
      After calibration according to the manufacturer’s instructions, the intaglio surfaces of the prostheses were digitized with an intraoral scanner (i500; Medit). A laboratory scanner was not used because of its limited scan angle and focal distance to the small area of the specimens.
      • Cho J.H.
      • Yoon H.I.
      • Han J.S.
      • Kim D.J.
      Trueness of the inner surface of monolithic crowns fabricated by milling of a fully sintered (Y, Nb)-TZP block in Chairside CAD-CAM system for single-visit dentistry.
      ,
      • Kim Y.K.
      • Han J.S.
      • Yoon H.I.
      Evaluation of intaglio surface trueness, wear, and fracture resistance of zirconia crown under simulated mastication: a comparative analysis between subtractive and additive manufacturing.
      ,
      • Kim Y.K.
      • Yoon H.I.
      • Kim D.J.
      • Han J.S.
      Comparative analysis on intaglio surface trueness, wear volume loss of antagonist, and fracture resistance of full-contour monolithic zirconia crown for single-visit dentistry under simulated mastication.
      The scan data of the intaglio surface of each prosthesis were superimposed on the identical field of the CAD prosthesis (reference CAD #1) and analyzed 3-dimensionally with a software program (Geomagic Control X; Geomagic Inc) to evaluate accuracy.
      • Osman R.B.
      • van der Veen A.J.
      • Huiberts D.
      • Wismeijer D.
      • Alharbi N.
      3D-printing zirconia implants; a dream or a reality? An in-vitro study evaluating the dimensional accuracy, surface topography and mechanical properties of printed zirconia implant and discs.
      ,
      • Wang W.
      • Yu H.
      • Liu Y.
      • Jiang X.
      • Gao B.
      Trueness analysis of zirconia crowns fabricated with 3-dimensional printing.
      ,
      • Alharbi N.
      • Osman R.B.
      • Wismeijer D.
      Factors influencing the dimensional accuracy of 3D-printed full-coverage dental restorations using stereolithography technology.
      With the best-fit alignment using an iterative closest point algorithm based on point-to-point distance measurement, the root mean square (RMS) values between the scan data and reference data were measured in μm to report the trueness of the prosthesis in 4 different regions of inspection: the intaglio surface and the occlusal, margin, and axial areas. Deviation maps were also displayed with a nominal deviation of ±50 μm and a critical deviation of ±500 μm.
      The MSA with the lowest RMS value for the single-unit prosthesis analysis was used to fabricate the specimens with the reference CAD #2. Anatomic contour 2-unit prostheses (n=10, Group DLP) were manufactured with the same DLP printer and zirconia suspension. All the printing parameters, postprocessing, and thermal treatment schedules were identical to those of the single-unit prosthesis. As a control group using the same reference CAD #2, 10 zirconia-based prostheses (Group MIL) were also fabricated from a milling block (Luxen ML Multi A3; Dentalmax Co) using a CAM software program (HyperDent; Follow-me! Technology) and a 5-axis milling machine (Arum 5X-300; Doowon). The milled prostheses were sintered according to the manufacturer’s protocol. For both groups, no further adjustments were made for the intaglio or outer surfaces after production. The intaglio surfaces of all 2-unit prostheses (DLP-generated and milled) were scanned with an intraoral scanner (i500; Medit). Scan data of the intaglio surfaces of the prostheses were superimposed on the identical field of the CAD prosthesis (reference CAD #2) and analyzed (Geomagic Control X; Geomagic Inc). With the best-fit alignment, the RMS values of the intaglio surface were measured to compare the trueness of the prosthesis in 4 different regions: the intaglio surface and occlusal, margin, and axial areas.
      Descriptive statistics using means and standard deviations, calculated from the measured RMS values at 4 different regions of inspection for each design, were analyzed for the single-unit and 2-unit prostheses. The normality and equality of variances were assessed by using the Shapiro–Wilk and Levene tests. Individual 1-Way ANOVAs were conducted to compare the RMS values of experimental MSA groups for each region of inspection, and a post hoc pairwise comparison was adjusted by using the Bonferroni method (α=.05). To compare the RMS values of the DLP and MIL groups, individual independent t tests were performed for each of the areas (α=.05). All data were analyzed by using a statistical software program (IBM SPSS Statistics, v25.0; IBM Corp) (α=.05).

      Results

      Of the 4 groups with different MSA values, all the specimens from 50-degree, 55-degree, and 60-degree MSA groups were successfully manufactured (Fig. 2). In contrast, all the specimens from the 65-degree MSA group were excluded from the analysis because of structural failures (delamination between layers or detachment from the platform) during AM. As shown in Figure 2, smaller MSA values represented more support structures spread over more expansive areas. Only the mean RMS values of the 50-degree MSA group were lower than 50 μm, regardless of the inspected regions (Fig. 3). The mean RMS value of the total area of the intaglio surface was significantly higher in the 60-degree MSA group than in the other groups (P<.001) (Table 2). At the margin area, the 60-degree MSA group showed a significantly higher RMS value than the other groups (P<.001). At the axial and occlusal areas, the 50-degree MSA group showed significantly lower RMS values than the other groups (P<.05). The 50-degree MSA group showed the highest trueness of the monolithic single-unit prosthesis (Fig. 4).
      Figure thumbnail gr2
      Figure 2Computer-aided design images of support structures incorporating MSA for anatomic contour dental prostheses successfully manufactured by digital light processing and zirconia suspension. A, Single-unit prosthesis with 50-degree MSA. B, Single-unit prosthesis with 55-degree MSA. C, Single-unit prosthesis with 60-degree MSA. D, Two-unit prosthesis with 50-degree MSA. MSA, maximum support attachment angle.
      Figure thumbnail gr3
      Figure 3Box-whisker plot of RMS (μm) values of anatomic contour zirconia single-unit prostheses measured at 4 different intaglio regions of inspection (margin, axial, occlusal, and intaglio surface areas) for 3 successfully produced groups of MSA (50, 55, and 60 degrees). Clinically acceptable limit of intaglio surface trueness (RMS) was suggested as 50 μm (dotted red line). MSA, maximum support attachment angle; RMS, root mean square.
      Table 2Mean ±standard deviation of root-mean-square (μm) values of anatomic contour zirconia single-unit prostheses measured at 4 different regions of inspection (margin, axial, occlusal, and intaglio surface areas) for 3 successfully produced groups of MSA
      GroupMargin AreaAxial AreaOcclusal AreaIntaglio Surface
      50-degree MSA18.3 ±2.1a14.5 ±3.6c23.9 ±2.5e35.0 ±6.8h
      55-degree MSA24.6 ±4.4a22.3 ±5.2d46.9 ±4.2f103.3 ±10.4i
      60-degree MSA36.2 ±9.4b24.8 ±6.3d65.1 ±15.1g138.0 ±24.0j
      MSA, maximum support attachment angle.
      Different letters within each column indicate statistically significant differences (P<.05).
      Figure thumbnail gr4
      Figure 4Three-dimensional color deviation map of intaglio surface of anatomic contour zirconia single-unit prostheses manufactured by digital light processing with 50-degree MSA. A, B, Intaglio surface (total). C, Margin area. D, Occlusal area. Nominal deviation ±50 μm, and critical deviation ±500 μm. MSA, maximum support attachment angle.
      An analysis of the DLP group with the 2-unit prosthesis specimens produced with an MSA of 50 degrees revealed statistically significant differences between the 2 prosthesis groups at each region of inspection (P<.001) (Table 3 and Fig. 5). The DLP group had significantly higher RMS values than the MIL group (P<.001); however, all specimens showed RMS values smaller than 50 μm, within the clinically acceptable range regardless of the inspection region (Figs. 5 and 6).
      Table 3Mean ±standard deviation and (minimum, maximum) of root-mean-square (μm) values of anatomic contour zirconia 2-unit prostheses produced with 50-degree MSA, measured at 4 different intaglio regions of inspection (intaglio surface and margin, axial, and occlusal areas)
      GroupIntaglio SurfaceMargin AreaAxial AreaOcclusal Area
      MIL19.1 ±1.3a(17.6, 21.2)23.5 ±4.3c(19.3, 31.5)16.9 ±1.9e(13.4, 19.5)15.1 ±2.3g(12.7, 20.4)
      DLP29.5 ±2.8b(25.9, 33.9)34.8 ±2.2d(31.8, 37.7)29.9 ±5.6f(23.8, 38.6)19.2 ±1.1h(17.3, 21.2)
      DLP, digital light processing; MIL, 5-axis milling; MSA, maximum support attachment angle.
      Different letters within each column indicate statistically significant differences (P<.001) with individual independent t tests.
      Figure thumbnail gr5
      Figure 5Box-whisker plot of RMS (μm) values of anatomic contour zirconia 2-unit prostheses produced with 50-degree MSA, measured at 4 different regions of inspection (intaglio surface, margin, axial, and occlusal areas) for groups MIL and DLP. Clinically acceptable limit of intaglio surface trueness (RMS) suggested as 50 μm (dotted red line). DLP, digital light processing; MIL, 5-axis milling; MSA, maximum support attachment angle; RMS, root-mean-square.
      Figure thumbnail gr6
      Figure 6Three-dimensional color deviation map of intaglio surface of anatomic contour zirconia two-unit prostheses manufactured by digital light processing (upper row) with 50-degree MSA and by five-axis milling (lower row). A, D, Intaglio surface. B, E, Margin area. C, F, Occlusal area. Nominal deviation ±50 μm, and critical deviation ±500 μm. MSA, maximum support attachment angle.

      Discussion

      Based on the results of this study, MSA significantly affected the intaglio surface trueness of DLP-generated anatomic contour dental prostheses, supporting the research hypothesis. Under the 50-degree MSA, the anatomic contour 2-unit prostheses produced by DLP showed mean RMS values less than 50 μm, which, even though significant differences were observed between the MIL and DLP groups, were clinically acceptable.
      Support structures are essential for successfully layering complex objects with sufficient detail and integrity during the AM process.
      • Baumgartner S.
      • Gmeiner R.
      • Schönherr J.A.
      • Stampfl J.
      Stereolithography-based additive manufacturing of lithium disilicate glass ceramic for dental applications.
      ,
      • Li H.
      • Song L.
      • Sun J.
      • Ma J.
      • Shen Z.
      Dental ceramic prostheses by stereolithography-based additive manufacturing: potentials and challenges.
      ,
      • Schönherr J.A.
      • Baumgartner S.
      • Hartmann M.
      • Stampfl J.
      Stereolithographic additive manufacturing of high precision glass ceramic parts.
      However, if the support structures are inadequately located on the AM objects, the increased duration of the printing process, extra work of postprocessing, and increased risk of damaging products can occur.
      • Baumgartner S.
      • Gmeiner R.
      • Schönherr J.A.
      • Stampfl J.
      Stereolithography-based additive manufacturing of lithium disilicate glass ceramic for dental applications.
      ,
      • Lian Q.
      • Wu X.
      • Li D.
      • et al.
      Accurate printing of a zirconia molar crown bridge using three-part auxiliary supports and ceramic mask projection stereolithography.
      • Li H.
      • Song L.
      • Sun J.
      • Ma J.
      • Shen Z.
      Dental ceramic prostheses by stereolithography-based additive manufacturing: potentials and challenges.
      • Schönherr J.A.
      • Baumgartner S.
      • Hartmann M.
      • Stampfl J.
      Stereolithographic additive manufacturing of high precision glass ceramic parts.
      • Revilla-León M.
      • Meyer M.J.
      • Zandinejad A.
      • Özcan M.
      Additive manufacturing technologies for processing zirconia in dental applications.
      ,
      • Piedra-Cascón W.
      • Krishnamurthy V.R.
      • Att W.
      • Revilla-León M.
      3D printing parameters, supporting structures, slicing, and post-processing procedures of vat-polymerization additive manufacturing technologies: a narrative review.
      Regarding the support attachment for the AM, some areas were theoretically considered suitable for the precision of the prosthesis, such as the proximal or smooth surfaces.
      • Schönherr J.A.
      • Baumgartner S.
      • Hartmann M.
      • Stampfl J.
      Stereolithographic additive manufacturing of high precision glass ceramic parts.
      However, support structures were not attached to those areas because they may have had thin-walled structures close to the margin areas, possibly damaging or deforming the AM object during postprocessing.
      • Lian Q.
      • Wu X.
      • Li D.
      • et al.
      Accurate printing of a zirconia molar crown bridge using three-part auxiliary supports and ceramic mask projection stereolithography.
      Considering the printability of zirconia suspension and its postprocessing, an MSA larger than 70 degrees or smaller than 45 degrees was regarded as inappropriate because it did not sufficiently support the entire structure, eventually leading to delamination or detachment by the movement of the recoating blade owing to insufficient adhesion to the build platform.
      • Li H.
      • Song L.
      • Sun J.
      • Ma J.
      • Shen Z.
      Dental ceramic prostheses by stereolithography-based additive manufacturing: potentials and challenges.
      The difficulty in removing support structures may be attributed to using the same material for the AM objects and support structures.
      • Li H.
      • Song L.
      • Sun J.
      • Ma J.
      • Shen Z.
      Dental ceramic prostheses by stereolithography-based additive manufacturing: potentials and challenges.
      During removal and further sintering, suboptimal support structures may lead to increased material loss and warpage during treatment, eventually leading to inaccurate geometry.
      • Schönherr J.A.
      • Baumgartner S.
      • Hartmann M.
      • Stampfl J.
      Stereolithographic additive manufacturing of high precision glass ceramic parts.
      The smaller MSA group had more support structures, which had to be removed during postprocessing, over more expansive areas than the larger MSA group. Nevertheless, more support structures are advantageous since the 50-degree MSA exhibited the highest trueness with mean RMS values less than 50 μm.
      The trueness of ceramic AM prostheses or fixtures has been evaluated.
      • Li R.
      • Xu T.
      • Wang Y.
      • Sun Y.
      Accuracy of zirconia crowns manufactured by stereolithography with an occlusal full-supporting structure: An in vitro study.
      • Osman R.B.
      • van der Veen A.J.
      • Huiberts D.
      • Wismeijer D.
      • Alharbi N.
      3D-printing zirconia implants; a dream or a reality? An in-vitro study evaluating the dimensional accuracy, surface topography and mechanical properties of printed zirconia implant and discs.
      • Li R.
      • Wang Y.
      • Hu M.
      • et al.
      Strength and adaptation of stereolithography-fabricated zirconia dental crowns: an in vitro study.
      • Revilla-León M.
      • Methani M.M.
      • Morton D.
      • Zandinejad A.
      Internal and marginal discrepancies associated with stereolithography (SLA) additively manufactured zirconia crowns.
      • Wang W.
      • Sun J.
      Dimensional accuracy and clinical adaptation of ceramic crowns fabricated with the stereolithography technique.
      • Wang W.
      • Yu H.
      • Liu Y.
      • Jiang X.
      • Gao B.
      Trueness analysis of zirconia crowns fabricated with 3-dimensional printing.
      The intaglio surface trueness of AM zirconia dental prostheses has been reported to be comparable with or inferior to those of milled prostheses; however, all the specimens were within the limit of clinical acceptance, with RMS values less than 50 μm.
      • Li R.
      • Xu T.
      • Wang Y.
      • Sun Y.
      Accuracy of zirconia crowns manufactured by stereolithography with an occlusal full-supporting structure: An in vitro study.
      ,
      • Li R.
      • Wang Y.
      • Hu M.
      • et al.
      Strength and adaptation of stereolithography-fabricated zirconia dental crowns: an in vitro study.
      • Revilla-León M.
      • Methani M.M.
      • Morton D.
      • Zandinejad A.
      Internal and marginal discrepancies associated with stereolithography (SLA) additively manufactured zirconia crowns.
      • Wang W.
      • Sun J.
      Dimensional accuracy and clinical adaptation of ceramic crowns fabricated with the stereolithography technique.
      • Wang W.
      • Yu H.
      • Liu Y.
      • Jiang X.
      • Gao B.
      Trueness analysis of zirconia crowns fabricated with 3-dimensional printing.
      The clinically acceptable limit in this study for trueness analysis was set to 50 μm as in previous studies.
      • Wang W.
      • Sun J.
      Dimensional accuracy and clinical adaptation of ceramic crowns fabricated with the stereolithography technique.
      ,
      • Peters M.C.
      • Delong R.
      • Pintado M.R.
      • Pallesen U.
      • Qvist V.
      • Douglas W.H.
      Comparison of two measurement techniques for clinical wear.
      • Cho J.H.
      • Yoon H.I.
      • Han J.S.
      • Kim D.J.
      Trueness of the inner surface of monolithic crowns fabricated by milling of a fully sintered (Y, Nb)-TZP block in Chairside CAD-CAM system for single-visit dentistry.
      • Rungrojwittayakul O.
      • Kan J.Y.
      • Shiozaki K.
      • et al.
      Accuracy of 3D Printed Models Created by Two Technologies of Printers with Different Designs of Model Base.
      • Al Hamad K.Q.
      • Al-Rashdan B.A.
      • Ayyad J.Q.
      • et al.
      Additive manufacturing of dental ceramics: a systematic review and meta-analysis.
      • Kim Y.K.
      • Han J.S.
      • Yoon H.I.
      Evaluation of intaglio surface trueness, wear, and fracture resistance of zirconia crown under simulated mastication: a comparative analysis between subtractive and additive manufacturing.
      Since the experiment with single-unit prostheses was to investigate the influence of MSA on trueness to meet the clinically acceptable guidelines, the absolute value of 50 μm was used to evaluate DLP-generated specimens rather than comparing DLP-generated specimens with a control group. In this study, using the 50-degree MSA, all the measured RMS values of the intaglio surfaces of DLP-generated zirconia single-unit and 2-unit prostheses were below 50 μm. The light source, techniques, and suspension composition could be different among the evaluated protocols; however, it can be concluded that AM enables the fabrication of anatomic contour monolithic prostheses with sufficient accuracy and reliability.
      • Khanlar L.N.
      • Salazar Rios A.
      • Tahmaseb A.
      • Zandinejad A.
      Additive manufacturing of zirconia ceramic and its application in clinical dentistry: a review.
      Optimizing the AM process for the material, considering the geometry and characteristics of the printed objects, is essential.
      • Piedra-Cascón W.
      • Krishnamurthy V.R.
      • Att W.
      • Revilla-León M.
      3D printing parameters, supporting structures, slicing, and post-processing procedures of vat-polymerization additive manufacturing technologies: a narrative review.
      In the present study, a top-down approach was used for DLP. A bottom-up approach may be useful because it requires fewer materials and is less expensive than a top-down approach.
      • Khanlar L.N.
      • Salazar Rios A.
      • Tahmaseb A.
      • Zandinejad A.
      Additive manufacturing of zirconia ceramic and its application in clinical dentistry: a review.
      ,
      • Komissarenko D.A.
      • Sokolov P.S.
      • Evstigneeva A.D.
      • Shmeleva I.A.
      • Dosovitsky A.E.
      Rheological and curing behavior of acrylate-based suspensions for the DLP 3D printing of complex zirconia parts.
      However, the top-down approach has advantages over the bottom-up approach for ceramic AM in terms of the number of support structures required, the light intensity without reduction, and no detachment movement at each layer.
      • Santoliquido O.
      • Colombo P.
      • Ortona A.
      Additive manufacturing of ceramic components by digital light processing: a comparison between the “bottom-up” and the “top-down” approaches.
      Limitations of the present study include that a dedicated device for zirconia AM has not been optimized for dental application. Among numerous AM techniques, none has been determined as the best option for producing ceramic dental prostheses. However, the effect of MSA on the trueness of anatomic contour prosthesis was revealed, highlighting the importance of support attachment design during AM with zirconia suspension. Only 2 types of dental prostheses (single- and 2-unit) were evaluated. The dimension and geometric complexity of various forms of fixed prostheses should be tested in the future. The sintering shrinkage and color reproduction of ceramic AM objects still have to be overcome for dental application. Lastly, the external surface of the prosthesis was not examined since the support structures were attached to the occlusal surface. The area of attached support structures might differ from the original design of the object after postprocessing, which necessitates further research.

      Conclusions

      Based on the findings of this in vitro study, the following conclusions were drawn:
      • 1.
        The intaglio surface trueness of anatomic contour monolithic prostheses can be improved by changing the maximum support attachment angle (MSA) during DLP using a zirconia suspension.
      • 2.
        With the use of an MSA of 50 degrees, the intaglio surface trueness of a DLP-generated prosthesis was less than 50 μm, which was within the clinically acceptable limit.

      CRediT authorship contribution statement

      Jun-Ho Cho: Writing - original draft. Hyung-In Yoon: Conceptualization, Data curation, Formal analysis, Supervision, Writing - review & editing. Jin-Ho Oh: Investigation, Methodology. Do-Hyun Kim: Resources, Software, Validation.

      References

        • Baumgartner S.
        • Gmeiner R.
        • Schönherr J.A.
        • Stampfl J.
        Stereolithography-based additive manufacturing of lithium disilicate glass ceramic for dental applications.
        Mater Sci Eng C Mater Biol Appl. 2020; 116111180
        • Lebon N.
        • Tapie L.
        • Duret F.
        • Attal J.-P.
        Understanding dental CAD/CAM for restorations-dental milling machines from a mechanical engineering viewpoint. Part A: chairside milling machines.
        Int J Comput Dent. 2016; 19: 45-62
        • Lian Q.
        • Wu X.
        • Li D.
        • et al.
        Accurate printing of a zirconia molar crown bridge using three-part auxiliary supports and ceramic mask projection stereolithography.
        Ceram Int. 2019; 45: 18814-18822
        • Li H.
        • Song L.
        • Sun J.
        • Ma J.
        • Shen Z.
        Dental ceramic prostheses by stereolithography-based additive manufacturing: potentials and challenges.
        Adv Appl Ceram. 2018; 118: 30-36
        • Schönherr J.A.
        • Baumgartner S.
        • Hartmann M.
        • Stampfl J.
        Stereolithographic additive manufacturing of high precision glass ceramic parts.
        Materials. 2020; 13: 1492
        • Revilla-León M.
        • Meyer M.J.
        • Zandinejad A.
        • Özcan M.
        Additive manufacturing technologies for processing zirconia in dental applications.
        Int J Comput Dent. 2020; 23: 27-37
        • Li R.
        • Xu T.
        • Wang Y.
        • Sun Y.
        Accuracy of zirconia crowns manufactured by stereolithography with an occlusal full-supporting structure: An in vitro study.
        J Prosthet Dent. 2022; ([Epub ahead of print.])https://doi.org/10.1016/j.prosdent.2022.01.015
        • Osman R.B.
        • van der Veen A.J.
        • Huiberts D.
        • Wismeijer D.
        • Alharbi N.
        3D-printing zirconia implants; a dream or a reality? An in-vitro study evaluating the dimensional accuracy, surface topography and mechanical properties of printed zirconia implant and discs.
        J Mech Behav Biomed Mater. 2017; 75: 521-528
        • Li R.
        • Wang Y.
        • Hu M.
        • et al.
        Strength and adaptation of stereolithography-fabricated zirconia dental crowns: an in vitro study.
        Int J Prosthodont. 2019; 32: 439-443
        • Revilla-León M.
        • Methani M.M.
        • Morton D.
        • Zandinejad A.
        Internal and marginal discrepancies associated with stereolithography (SLA) additively manufactured zirconia crowns.
        J Prosthet Dent. 2020; 124: 730-737
        • Wang W.
        • Sun J.
        Dimensional accuracy and clinical adaptation of ceramic crowns fabricated with the stereolithography technique.
        J Prosthet Dent. 2021; 125: 657-663
        • Wang W.
        • Yu H.
        • Liu Y.
        • Jiang X.
        • Gao B.
        Trueness analysis of zirconia crowns fabricated with 3-dimensional printing.
        J Prosthet Dent. 2019; 121: 285-291
        • Griffith M.L.
        • Halloran J.W.
        Freeform fabrication of ceramics via stereolithography.
        J Am Ceram Soc. 1996; 79: 2601-2608
        • Khanlar L.N.
        • Salazar Rios A.
        • Tahmaseb A.
        • Zandinejad A.
        Additive manufacturing of zirconia ceramic and its application in clinical dentistry: a review.
        Dent J. 2021; 9: 104
        • Piedra-Cascón W.
        • Krishnamurthy V.R.
        • Att W.
        • Revilla-León M.
        3D printing parameters, supporting structures, slicing, and post-processing procedures of vat-polymerization additive manufacturing technologies: a narrative review.
        J Dent. 2021; 109: 103630
        • Lee M.P.
        • Cooper G.J.
        • Hinkley T.
        • Gibson G.M.
        • Padgett M.J.
        • Cronin L.
        Development of a 3D printer using scanning projection stereolithography.
        Sci Rep. 2015; 5: 1-5
        • Silva N.R.
        • Witek L.
        • Coelho P.G.
        • Thompson V.P.
        • Rekow E.D.
        • Smay J.
        Additive CAD/CAM process for dental prostheses.
        J Prosthodont. 2011; 20: 93-96
        • Alharbi N.
        • Osman R.B.
        • Wismeijer D.
        Factors influencing the dimensional accuracy of 3D-printed full-coverage dental restorations using stereolithography technology.
        Int J Prosthodont. 2016; 29: 503-510
        • Braian M.
        • Jimbo R.
        • Wennerberg A.
        Production tolerance of additive manufactured polymeric objects for clinical applications.
        Dent Mater. 2016; 32: 853-861
        • Peters M.C.
        • Delong R.
        • Pintado M.R.
        • Pallesen U.
        • Qvist V.
        • Douglas W.H.
        Comparison of two measurement techniques for clinical wear.
        J Dent. 1999; 27: 479-485
        • Cho J.H.
        • Yoon H.I.
        • Han J.S.
        • Kim D.J.
        Trueness of the inner surface of monolithic crowns fabricated by milling of a fully sintered (Y, Nb)-TZP block in Chairside CAD-CAM system for single-visit dentistry.
        Materials. 2019; 12: 3253
        • Rungrojwittayakul O.
        • Kan J.Y.
        • Shiozaki K.
        • et al.
        Accuracy of 3D Printed Models Created by Two Technologies of Printers with Different Designs of Model Base.
        J Prosthodont. 2020; 29: 124-128
        • Al Hamad K.Q.
        • Al-Rashdan B.A.
        • Ayyad J.Q.
        • et al.
        Additive manufacturing of dental ceramics: a systematic review and meta-analysis.
        J Prosthodont. 2022; 31: e67-e86
        • Kim Y.K.
        • Han J.S.
        • Yoon H.I.
        Evaluation of intaglio surface trueness, wear, and fracture resistance of zirconia crown under simulated mastication: a comparative analysis between subtractive and additive manufacturing.
        J Adv Prosthodont. 2022; 14: 122-132
        • Kim Y.K.
        • Yoon H.I.
        • Kim D.J.
        • Han J.S.
        Comparative analysis on intaglio surface trueness, wear volume loss of antagonist, and fracture resistance of full-contour monolithic zirconia crown for single-visit dentistry under simulated mastication.
        J Adv Prosthodont. 2022; 14: 173-181
        • Komissarenko D.A.
        • Sokolov P.S.
        • Evstigneeva A.D.
        • Shmeleva I.A.
        • Dosovitsky A.E.
        Rheological and curing behavior of acrylate-based suspensions for the DLP 3D printing of complex zirconia parts.
        Materials. 2018; 11: 2350
        • Santoliquido O.
        • Colombo P.
        • Ortona A.
        Additive manufacturing of ceramic components by digital light processing: a comparison between the “bottom-up” and the “top-down” approaches.
        J Eur Ceram Soc. 2019; 39: 2140-2148