Osseointegration in relation to drilling speed in the preparation of dental implants sites: A systematic review

      Abstract

      Statement of problem

      The drilling speed used for preparing dental implants may affect bone-implant contact (BIC), implant stability quotient (ISQ), and bone area fraction occupancy (BAFO). Different rotational speeds and the presence or absence of irrigation during site preparation have been investigated, but an established protocol for achieving the best osseointegration results is lacking.

      Purpose

      The purpose of this systematic review was to investigate the influence of drill rotational speed on bone drilling for dental implant placement and its relationship with osseointegration.

      Material and methods

      This review included the preferred reporting items for systematic reviews and meta-analyses (PRISMA) and was registered in the international prospective register of systematic reviews (PROSPERO) database. Electronic searches were performed in the MEDLINE (PubMed), Scopus, Science Direct, and Embase databases. The risk of bias was analyzed by using the systematic review center for laboratory animal experimentation (SYRCLE).

      Results

      A total of 1282 articles were found, and after removing duplicates and applying the eligibility criteria to in vivo articles on animals that addressed drilling speed and its relationship to osseointegration, 8 articles were selected for analysis. Of these, 5 articles showed no statistical differences, and 3 others showed significantly better osseointegration results by analyzing the parameters of BIC, BAFO, ISQs, and pull-out forces (PoFs). In all selected articles, high-speed drilling was performed with irrigation.

      Conclusions

      Although drilling speed seems to affect bone perforation, no definitive protocol was found in the literature consulted. The results vary depending on the combination of different factors, including bone type, irrigation, and drilling speed.
      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Journal of Prosthetic Dentistry
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Javed F.
        • Romanos G.E.
        The role of primary stability for successful immediate loading of dental implants. A literature review.
        J Dent. 2010; 38: 612-620
        • Witek L.
        • Parra M.
        • Tovar N.
        • et al.
        Effect of surgical instrumentation variables on the osseointegration of narrow- and wide-diameter short implants.
        J Oral Maxillofac Surg. 2021; 79: 346-355
        • Valente M.L.C.
        • Castro D.T.
        • Shimano A.C.
        • Lepri C.P.
        • Reis A.C.
        Analyzing the influence of a new dental implant design on primary stability.
        Clin Implant Dent Relat Res. 2016; 18: 168-173
        • Santos M.V.
        • Elias C.N.
        • Cavalcanti L.J.H.
        The effects of superficial roughness and design on the primary stability of dental implants.
        Clin Implant Dent Relat Res. 2011; 3: 215-223
        • Yeniyol S.
        • Jimbo R.
        • Marin C.
        • Tovar N.
        • Janal M.N.
        • Coelho P.G.
        The effect of drilling speed on early bone healing to oral implants.
        Oral Surg Oral Med Oral Pathol Oral Radiol. 2013; 116: 550-555
        • Eliasa C.N.
        • Rocha F.A.
        • Nascimento A.L.
        • Coelho P.G.
        Influence of implant shape, surface morphology, surgical technique and bone quality on the primary stability of dental implants.
        J Mech Behav Biomed Mater. 2012; 16: 169-180
        • Li Z.
        • Arioka M.
        • Liu Y.
        • et al.
        Effects of condensation and compressive strain on implant primary stability: a longitudinal, in vivo, multiscale study in mice.
        Bone Joint Res. 2020; 9: 60-70
        • Marquezan M.
        • Osório A.
        • Sant’Anna E.
        • Souza M.M.
        • Maia L.
        Does bone mineral density influence the primary stability of dental implants? A systematic review.
        Clin Oral Implants Res. 2012; 7: 767-774
        • Turkyilmaz I.
        • Aksoy U.
        • McGlumphy E.A.
        Two alternative surgical techniques for enhancing primary implant stability in the posterior maxilla: a clinical study including bone density, insertion torque, and resonance frequency analysis data.
        Clin Implant Dent Relat Res. 2008; 4: 231-237
        • Capparé P.
        • Vinci R.
        • di Stefano D.A.
        • et al.
        Correlation between initial BIC and the insertion torque/depth integral recorded with an instantaneous torque-measuring implant motor: an in vivo study.
        Clin Implant Dent Relat Res. 2015; 17: 613-620
        • Ottoni J.M.P.
        • Oliveira Z.F.L.
        • Mansini R.
        • Cabral A.M.
        Correlation between placement torque and survival of single-tooth implants.
        Int J Oral Maxillofac Implants. 2005; 20: 1-9
        • Hawais S.
        Enhancing implant stability with osseodensification- A case report with 2-year follow-up.
        Implant practice. 2014; 8: 1-5
        • Huwais S.
        • Meyer E.
        A novel osseous densification approach in implant osteotomy preparation to increase biomechanical primary stability, bone mineral density, and bone-to-implant contact.
        Int J Oral Maxillofac Implants. 2017; 1: 27-36
        • Favero V.
        • Sakuma S.
        • Alccayhuaman K.A.A.
        • Benedetto G.A.
        • Bengazi F.
        • Botticelli D.
        Healing at sites prepared using different drilling protocols. An experimental study in the tibiae of sheep.
        PLoS One. 2018; 13: 1-19
        • Mirzaie T.
        • Rouhi G.
        • Mehdi Dehghan M.
        • Farzad-Mohajeri S.
        • Barikani H.
        Dental implants’ stability dependence on rotational speed and feed-rate of drilling: In-vivo and ex-vivo investigations.
        J Biomech. 2021; 11: 1-7
        • Geckili O.
        • Bilhan H.
        • Geckili E.
        • Barca-Dayan E.
        • Dayan C.
        • Bural C.
        Is clinical experience important for obtaining the primary stability of dental implants with aggressive threads? An ex vivo study.
        Med Oral Patol Oral Cir Bucal. 2019; 24: 254-259
        • Planinić D.
        • Dubravica I.
        • Šarac Z.
        • et al.
        Comparison of different surgical procedures on the stability of dental implants in posterior maxilla: a randomized clinical study.
        J Stomatol Oral Maxillofac Surg. 2021; 122: 487-493
        • Tardelli J.D.C.
        • Valente M.L.C.
        • Macedo A.P.
        • Reis A.C.
        Evaluation of biomechanical and stress distribution of different dental implant designs: primary stability and photoelastic analysis.
        IRBM. 2022; 43: 100-106
        • Giro G.
        • Marin C.
        • Granato R.
        • et al.
        Effect of drilling technique on the early integration of plateau root form endosteal implants: an experimental study in dogs.
        J Oral Maxillofac Surg. 2011; 69: 2158-2163
        • Holm-Pedersen P.
        • Lang N.P.
        • Müller F.
        What are the longevities of teeth and oral implants?.
        Clin Oral Implants Res. 2007; 18: 15-19
        • Tardelli J.D.C.
        • Reis A.C.
        Influence of surface electric charge of Ti implants on osteoblastic interaction: a systematic review.
        Saudi Dent J. 2022; 34: 335-345
        • Brånemark R.
        • Brånemark P.
        • Rydevik B.
        • Myers R.R.
        Osseointegration in skeletal reconstruction and rehabilitation: a review and the VA San Diego Healthcare System.
        J Rehabil Res Dev. 2001; 38: 1-4
        • Trisi P.
        • Perfetti G.
        • Baldoni E.
        • Berardi D.
        • Colagiovanni M.
        • Scogna G.
        Implant micromotion is related to peak insertion torque and bone density.
        Clin Oral Implants Res. 2009; 5: 467-471
        • Albrektsson T.
        • Wennerberg A.
        On osseointegration in relation to implant surfaces.
        Clin Implant Dent Relat Res. 2019; 21: 4-7
        • Albrektsson T.
        • Johansson C.
        Osteoinduction, osteoconduction and osseointegration.
        Eur Spine J. 2001; 10: 96-101
        • Pai U.
        • Rodrigues S.
        • Talreja K.
        • Mundathaje M.
        Osseodensification – a novel approach in implant dentistry.
        J Indian Prosthodont Soc. 2018; 3: 196-200
        • Halldin A.
        • Jimbo R.
        • Johansson C.B.
        • et al.
        The effect of static bone strain on implant stability and bone remodeling.
        Bone. 2011; 49: 783-789
        • Lee J.E.
        • Rabin Y.
        • Ozdoganlar O.B.
        A new thermal model for bone drilling with applications to orthopedic surgery.
        Med Eng Phys. 2011; 33: 1234-1244
        • Carvalho A.C.
        • Queiroz T.P.
        • Okamoto R.
        • Margonar R.
        • Garcia Jr., I.R.
        • Magro Filho O.
        Evaluation of bone heating, immediate bone cell viability, and wear of high-resistance drills after the creation of implant osteotomies in rabbit tibias.
        Int J Oral Maxillofac Implants. 2011; 26: 1193-1201
        • Coyac B.R.
        • Salvi G.
        • Leahy B.
        • et al.
        A novel system exploits bone debris for implant osseointegration.
        J Periodontol. 2021; 92: 716-726
        • Bahat O.
        • Yin X.
        • Holst S.
        • et al.
        An osteotomy tool that preserves bone viability: evaluation in preclinical and clinical settings.
        J Clin Med. 2022; 11: 1-15
        • Cemil S.
        • Cansiz E.
        • Tanyel C.
        • Soluk M.
        • Selvi F.
        • Cebi Z.
        The effect of irrigation temperature on bone healing.
        Int J Med Sci. 2011; 8: 704-708
        • Eriksson R.A.
        • Adell R.
        Temperatures during drilling for the placement of implants using the osseointegration technique.
        J Oral Maxlllofac Surg. 1986; 44: 4-7
        • Tsiagadigui J.G.
        • Ndiwe B.
        • Ngo Yamben M.A.
        • Fotio N.
        • Belinga F.E.
        • Njeugna E.
        The effects of multiple drilling of a bone with the same drill bit: thermal and force analysis.
        Heliyon. 2022; 8
        • Augustin G.
        • Zigman T.
        • Davila S.
        • et al.
        Cortical bone drilling and thermal osteonecrosis.
        Clin Biomech. 2012; 27: 313-325
        • Kim S.J.
        • Yoo J.
        • Kim Y.S.
        • Shin S.W.
        Temperature change in pig rib bone during implant site preparation by low-speed drilling.
        J Appl Oral Sci. 2010; 18: 522-527
        • Tawy G.F.
        • Rowe P.J.
        • Riches P.E.
        Thermal damage done to bone by burring and sawing with and without irrigation in knee arthroplasty.
        J Arthroplasty. 2016; 31: 1102-1108
        • Dolan E.B.
        • Haugh M.G.
        • Tallon D.
        • Casey C.
        • McNamara L.M.
        Heat-shock-induced cellular responses to temperature elevations occurring during orthopaedic cutting.
        J R Soc Interface. 2012; 9: 3503-3513
        • Eriksson A.R.
        • Albrektsson T.
        Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit.
        J Prosthet Dent. 1983; 50: 101-107
        • Gaspar J.
        • Borrecho G.
        • Oliveira P.
        • Salvado F.
        • Martins J.
        • Santos D.
        Osteotomy at low-speed drilling without irrigation versus high-speed drilling with irrigation: an experimental study.
        Acta Med Port. 2013; 26: 231-236
        • Sharawy M.
        • Misch C.E.
        • Weller N.
        • Tehemar S.
        Heat generation during implant drilling: the significance of motor speed.
        J Oral Maxillofac Surg. 2002; 60: 1160-1169
        • Lee J.E.
        • Ozdoganlar O.B.
        • Rabin Y.
        An experimental investigation on thermal exposure during bone drilling.
        Med Eng Phys. 2012; 10: 1510-1520
        • Benington I.C.
        • Biagioni P.A.
        • Briggs J.
        • Sheridan S.
        • Lamey P.J.
        Thermal changes observed at implant sites during internal and external irrigation.
        Clin Oral Implants Res. 2002; 13: 293-297
        • Chen C.H.
        • Coyac B.R.
        • Arioka M.
        • et al.
        A novel osteotomy preparation technique to preserve implant site viability and enhance osteogenesis.
        J Clin Med. 2019; 8: 1-13
        • Ercoli C.
        • Funkenbusch P.D.
        • Lee H.J.
        • Moss M.E.
        • Graser G.N.
        The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants: a study of drill durability.
        Int J Oral Maxillofac Implants. 2004; 19: 1-16
        • Strbac G.D.
        • Giannis K.
        • Unger E.
        • et al.
        Drilling- and withdrawing-related thermal changes during implant site osteotomies.
        Clin Implant Dent Relat Res. 2015; 1: 32-43
        • Berman A.T.
        • Reid J.S.
        • Yanicko D.R.
        • Sih G.C.
        • Zimmerman M.R.
        Thermally induced bone necrosis in rabbits relation to implant failure in humans.
        Clin Orthop Relat Res. 1984; 186: 1-9
        • Eriksson A.R.
        • Albrektsson T.
        • Albrektsson B.
        Heat caused by drilling cortical bone: temperature measured in vivo in patients and animals.
        Acta Orthop. 1984; 55: 629-631
        • Harris B.H.
        • Kohles S.S.
        Effects of mechanical and thermal fatigue on dental drill performance.
        J Oral Maxillofac Implants. 2001; 16: 819-826
        • Misir A.F.
        • Sumer M.
        • Yenisey M.
        • Ergioglu E.
        Effect of surgical drill guide on heat generated from implant drilling.
        J Oral Maxillofac Surg. 2009; 12: 2663-2668
        • Seo D.U.
        • Kim S.G.
        • Oh J.S.
        • Lim S.C.
        Comparative study on early osseointegration of implants according to various drilling speeds in the mandible of dogs.
        Implant Dent. 2017; 26: 841-847
        • Can M.
        • Koluaçik S.
        • Bahçe E.
        • Gokce H.
        • Tecellioglu F.S.
        Investigation of thermal damage in bone drilling: Hybrid processing method and pathological evaluation of existing methods.
        J Mech Behav Biomed Mater. 2022; 126: 1-11
        • Ein-Afshar M.J.
        • Shahrezaee M.
        • Shahrezaee M.H.
        • Sharifzadeh S.R.
        Biomechanical evaluation of temperature rising and applied force in controlled cortical bone drilling an animal in vitro study.
        Arch Bone and Jt Surg. 2020; 8: 605-612
        • Alam K.
        • Piya S.
        • Al-Ghaithi A.
        • Silberschmidth V.
        Experimental investigation on the effect of drill quality on the performance of bone drilling.
        BioMed Tech. 2020; 65: 113-120
        • Lughmani W.A.
        • Bouazza-Marouf K.
        • Ashcroft I.
        Finite element modeling and experimentation of bone drilling forces.
        J Phys Conf Ser. 2013; 451: 1-8
        • Tuijthof G.J.M.
        • Frühwirt C.
        • Kment C.
        Influence of tool geometry on drilling performance of cortical and trabecular bone.
        Med Eng Phys. 2013; 35: 1165-1172
        • Sui J.
        • Sugita N.
        Experimental study of thrust force and torque for drilling cortical bone.
        Ann Biomed Eng. 2019; 47: 802-812
        • Koopaie M.
        • Kolahdouz S.
        • Kolahdouz E.M.
        Comparison of wear and temperature of zirconia and tungsten carbide tools in drilling bone: in vitro and finite element analysis.
        Br J Oral Maxillofac Surg. 2019; 57: 557-565
        • Fujiwara S.
        • Botticelli D.
        • Kaneko N.
        • Urbizo Velez J.
        • Tumedei M.
        • Bengazi F.
        Effect of low-speed drilling without irrigation on osseointegration: an experimental study in dogs.
        Oral Maxillofac Surg. 2021; 26: 1-7
        • Allan W.
        • Williams E.D.
        • Kerawala C.J.
        Effects of repeated drill use on temperature of bone during preparation for osteosynthesis self-tapping screws.
        Br J Oral Maxillofac Surg. 2005; 43: 314-319
        • Chacon G.E.
        • Bower D.L.
        • Larsen P.E.
        • McGlumphy E.A.
        • Beck F.M.
        Heat production by 3 implant drill systems after repeated drilling and sterilization.
        J Oral Maxillofac Surg. 2006; 64: 265-269
        • Koo K.T.
        • Kim M.H.
        • Kim H.Y.
        • Wikesjö U.M.E.
        • Yang J.H.
        • Yeo I.S.
        Effects of implant drill wear, irrigation, and drill materials on heat generation in osteotomy sites.
        J Oral Implantol. 2015; 41: 19-23
        • Marzook H.A.M.
        • Yousef E.A.
        • Denewar M.
        • Farahat M.R.L.
        In-vitro assessment of bone viability with different implant drill speeds.
        Br J Oral Maxillofac Surg. 2020; 58: 301-306
        • Witek L.
        • Alifarag A.M.
        • Tovar N.
        • et al.
        Osteogenic parameters surrounding trabecular tantalum metal implants in osteotomies prepared via osseodensification drilling.
        Med Oral Patol Oral Cir Bucal. 2019; 24: 764-769
        • Aghvami M.
        • Brunski J.B.
        • Tulu U.S.
        • Chen C.H.
        • Helms J.A.
        A thermal and biological analysis of bone drilling.
        J Biomech Eng. 2018; 140: 1-10
        • Coyac B.R.
        • Sun Q.
        • Leahy B.
        • et al.
        Optimizing autologous bone contribution to implant osseointegration.
        J Periodontol. 2020; 91: 1632-1644
        • Soriano J.
        • Garay A.
        • Aristimuño P.
        • Iriarte L.M.
        • Eguren J.A.
        • Arrazola P.J.
        Effects of rotational speed, feed rate and tool type on temperatures and cutting forces when drilling bovine cortical bone.
        Mach Sci Tecnol. 2013; 17: 611-636
        • Barrio R.A.L.
        • Paula W.N.
        • Spin-Neto R.
        • Souza J.A.C.
        • Oliveira G.J.P.L.
        • Marcantonio-Junior E.
        Effect of 2 different drilling speeds on the osseointegration of implants placed with flapless guided surgery: a study in rabbits.
        Implant Dent. 2017; 26: 882-887
        • Hooijmans C.R.
        • Rovers M.M.
        • Vries R.B.M.
        • Leenaars M.
        • Ritskes-Hoitinga M.
        • Langendam M.W.
        SYRCLE’s risk of bias tool for animal studies.
        BMC Med Res Methodol. 2014; 14: 1-9
        • Al Asfour A.A.
        • Abuzayeda M.
        • Murray C.A.
        • Kamal M.
        • Zafiropoulos G.G.
        Considerations regarding dental implant site preparation using osseo densification osteotomy. A critical review.
        Adv Dent Oral Health. 2020; 13: 1-8