Positional influence of center of masticatory forces on occlusal contact forces using a digital occlusal analyzer

Open AccessPublished:March 30, 2023DOI:https://doi.org/10.1016/j.prosdent.2023.03.007

      Abstract

      Statement of problem

      Digital occlusal analyzers allow the recording of dental contact forces. Some authors assume a unique location for the center of contact forces at the position of maximum intercuspation, while others indicate variations in dental contact forces when recorded at different times of the day. Which approach is more appropriate is unclear.

      Purpose

      The purpose of this in vitro study was to analyze whether a change in the balance of masticatory forces influences the location of the center of contact forces and its magnitude.

      Material and methods

      Three different dental casts, selected under dental criteria, were mounted in maximum intercuspation on a semiadjustable articulator equipped with a pattern indicating 9 different force application points (intersection point between 3 longitudinal rows and 3 transverse columns). A force of constant magnitude (169 N) was applied 10 times at each of the application points, and occlusal forces were recorded with a digital occlusal analyzer. Then, two variables were studied: the location of the center of contact forces and its magnitude. Each force application position (9 positions × 3 dental casts=27 in total) was repeated 10 times, and measured data were statistically analyzed with 2-way repeated measures ANOVA (α=.05) test.

      Results

      The repeatability of the method indicated that the coefficient of variation mean was 0.37% in the location of the center of contact forces and that its magnitude was 3.70%. The 2-way repeated measures ANOVA test revealed statistically significant variations in the location of the center of contact forces and its magnitude, revealing that longitudinal changes of the application point of masticatory forces affected the magnitude of contact forces and that longitudinal and transverse changes of the application point of masticatory forces affected the location of the center of contact forces.

      Conclusions

      The location of the center of contact force and its magnitude provided by a digital occlusal analyzer at the position of maximum intercuspation are not necessarily unique to each articulated dental cast. Even if the intensity of the masticatory force remains unchanged, changes in its lateral or longitudinal balance also influence the result of the occlusion forces.
      Clinical Implications
      Transverse equilibrium changes in masticatory forces can change occlusion measured with a digital occlusal analyzer. Dentists should consider this circumstance in their diagnoses, as fatigue or pain could change the masticatory force balance.
      Digital occlusal analyzers were developed to detect contacts between teeth and to measure the relative intensity of force at those contacts at each instant throughout occlusion.
      • Maness W.L.
      • Chapman R.J.
      • Dario L.D.
      Laboratory evaluation of a direct reading digital occlusal sensor.
      The first to be marketed and the most studied device (T-Scan; Tekscan Inc) has a measurable area unit of 1.27×1.27 mm, which corresponds to the area of a sensel.
      • Bozhkova T.P.
      The .T-SCAN system in evaluating occlusal contacts.
      This device, with its own software program (Fig. 1), processes measured data and provides qualitative graphical information on the intensity of force on each sensel and shows the location of the center of contact forces (CO-CF) together with its trajectory during occlusion.
      • Olivieri F.
      • Kang K.H.
      • Hirayama H.
      • Maness W.L.
      New method for analyzing complete denture occlusion using the center of force concept: a clinical report.
      ,
      • Shinogaya T.
      • Tanaka Y.
      • Toda S.
      • Hayakawa I.
      A new approach to evaluating occlusal support by analyzing the center of the bite force.
      Figure thumbnail gr1
      Figure 1Digital occlusal analyzer (T-Scan; Tekscan Inc) results screen with red and white rhombus indicating location of center of forces in maximum intercuspation position together with its trajectory during occlusion.
      Output data provided by the device, and especially the location of CO-CF regarding the mid-sagittal plane, are intended to help the dentist detect occlusal quality. These data are also used for other clinical purposes, such as to detect and correct diseases of the temporomandibular joint,
      • Ferrato G.
      Digital evaluation of occlusal forces: comparison between healthy subjects and TMD patients.
      ,
      • Li B.Y.
      • Zhou L.J.
      • Guo S.X.
      • Zhang Y.
      • Lu L.
      • Wang M.Q.
      An investigation on the simultaneously recorded occlusion contact and surface electromyographic activity for patients with unilateral temporomandibular disorders pain.
      to evaluate rehabilitations with or without implants,
      • Shinogaya T.
      • Tanaka Y.
      • Toda S.
      • Hayakawa I.
      A new approach to evaluating occlusal support by analyzing the center of the bite force.
      ,
      • Liu C.W.
      • Chang Y.M.
      • Shen Y.F.
      • Hong H.H.
      Using the T-scan III system to analyze occlusal function in mandibular reconstruction patients: a pilot study.
      to evaluate the outcome of orthodontic treatment,
      • Wieczorek A.
      • Loster J.E.
      Activity of the masticatory muscles and occlusal contacts in young adults with and without orthodontic treatment.
      and to address various dental problems such as bruxism,
      • Bandodkar S.
      • Tripathi S.
      • Chand P.
      • et al.
      A study to evaluate psychological and occlusal parameters in bruxism.
      severe tooth wear,
      • Sierpinska T.
      • Kuc J.
      • Golebiewska M.
      Assessment of masticatory muscle activity and occlusion time in patients with advanced tooth wear.
      and muscle pain.
      • Di Berardino F.
      • Filipponi E.
      • Schiappadori M.
      • Forti S.
      • Zanetti D.
      • Cesarani A.
      The occlusal imaging and analysis system by T-scan III in tinnitus patients.
      Despite providing useful information for dentists, the use of these devices has not been as widespread as initially expected,
      • Bozhkova T.
      • Musurlieva N.
      • Slavchev D.
      • Dimitrova M.
      Occlusal indicators used in dental practice: a Survey Study.
      possibly because of doubts about the reliability of their measurements. The reliability of contact positions provided by such a device has been widely analyzed and compared in relation to other methods,
      • Saraçoǧlu A.
      • Özpinar B.
      In vivo and in vitro evaluation of occlusal indicator sensitivity.
      • Bozhkova T.
      • Musurlieva N.
      • Slavchev D.
      Comparative study qualitative and quantitative techniques in the study of occlusion.
      • Bostancıoğlu S.E.
      • Toğay A.
      • Tamam E.
      Comparison of two different digital occlusal analysis methods.
      • Fraile C.
      • Ferreiroa A.
      • Solaberrieta GP E.
      Intraoral versus extraoral digital occlusal records: a pilot study.
      • Solaberrieta E.
      • Etxaniz O.
      • Otegi J.R.
      • Brizuela A.
      • Pradies G.
      Customized procedure to display T-Scan occlusal contacts.
      whereas the reliability of the contact force (CF) values it provides has been less studied.
      • Cerna M.
      • Ferreira R.
      • Zaror C.
      • Navarro P.
      • Sandoval P.
      Validity and reliability of the T-Scan III for measuring force under laboratory conditions.
      • Lee W.
      • Kwon H.B.
      • Kim M.J.
      • Lim Y.J.
      Determination of the reliability and repeatability of a quantitative occlusal analyzer by using a piezoelectric film sensor: an in vitro study.
      • Kerstein R.B.
      • Radke J.
      In-vitro consistency testing of the T-Scan 10 relative force measurement system in-vitro consistency testing of the T-Scan 10 relative force measurement system.
      Another difficulty with the clinical use of such a device has been the inability to interpret the data obtained. One such difficulty is the location of the CO-CF, which has been studied.
      • Olivieri F.
      • Kang K.H.
      • Hirayama H.
      • Maness W.L.
      New method for analyzing complete denture occlusion using the center of force concept: a clinical report.
      ,
      • Maness W.L.
      • Podoloff R.
      Distribution of occlusal contacts in maximum intercuspation.
      ,
      • Mizui M.
      • Nabeshima F.
      • Tosa J.
      • Tanaka M.
      • Kawazoe T.
      Quantitative analysis of occlusal balance in intercuspal position using the T-Scan system.
      While some authors assumed that each individual’s dentition has a unique location of CO-CF at maximum intercuspation position (MIP),
      • Maness W.L.
      • Podoloff R.
      Distribution of occlusal contacts in maximum intercuspation.
      • Mizui M.
      • Nabeshima F.
      • Tosa J.
      • Tanaka M.
      • Kawazoe T.
      Quantitative analysis of occlusal balance in intercuspal position using the T-Scan system.
      • Türp J.C.
      • Greene C.S.
      • Strub J.R.
      Dental occlusion: a critical reflection on past, present and future concepts.
      other authors concluded that occlusion cannot be considered static and unique in MIP and that intensities at contact points vary throughout the day.
      • Wiechens B.
      • Brockmeyer P.
      • Wassmann T.
      • Rödiger M.
      • Wiessner A.
      • Bürgers R.
      Time of day-dependent deviations in dynamic and static occlusion: a prospective clinical study.
      ,
      • Berry D.C.
      • Singh B.P.
      Daily variations in occlusal contacts.
      This last statement has been considered in the past, pointing to the level of contraction of the masticatory muscles as a possible cause of this variation.
      • Berry D.C.
      • Singh B.P.
      Daily variations in occlusal contacts.
      Three forces act simultaneously in the MIP of occlusion (Fig. 2): the forces exerted voluntarily by masticatory muscles, CFs between teeth, and forces on the temporomandibular joint. The sum of these 3 forces should be in equilibrium. Masticatory force and CF are different from each other, and each has a resultant that intersects the occlusal plane at a point called the center of forces; they also have a different force magnitude (Table 1).
      Figure thumbnail gr2
      Figure 2Force type that appears in occlusion when maximum intercuspation position has been reached.
      Table 1Abbreviations related to forces when maximum intercuspation position has been reached
      Occlusion Force TypeResultantMagnitudeCenter of Force
      Masticatory force: input force of test(MF)(MF-M)(CO-MF)
      Device-measured contact forces: output force of test(CF)(CF-M)(CO-CF)
      Temporomandibular joint forceNot usedNot usedNot used
      CF, contact force; CF-M, contact force magnitude; CO-CF, center of contact force; CO-MF, center of masticatory forces; MF, masticatory force; MF-M, magnitude of masticatory forces.
      When masticatory muscles exert compressive force, the maxillary and mandibular dentition come together until they reach the MIP, and, in response, the CFs, which are the forces measured by a digital occlusal analyzer, rise at all dental contacts. Reaction forces also occur at the condyles of the temporomandibular joint, although these are not recorded with these devices.
      Any physiological inequality, or any lack of lateral symmetry of the main masticatory muscles (masseter and temporalis muscle),
      • Raadsheer M.C.
      • Van Eijden T.M.G.J.
      • Van Ginkel F.C.
      • Prahl-Andersen B.
      Contribution of jaw muscle size and craniofacial morphology to human bite force magnitude.
      ,
      • Peck C.C.
      Biomechanics of occlusion - implications for oral rehabilitation.
      is considered to result in a noncoincidental location of the center of masticatory forces (CO-MF) in the mid-sagittal plane. In addition, each individual can voluntarily or involuntarily vary the intensity of muscle contraction.
      • Shimada A.
      • Baad-Hansen L.
      • Svensson P.
      Effect of experimental jaw muscle pain on dynamic bite force during mastication.
      ,
      • Goiato M.C.
      • Zuim P.R.J.
      • Moreno A.
      • et al.
      Does pain in the masseter and anterior temporal muscles influence maximal bite force?.
      As a consequence, the CO-MF position will move longitudinally in an anterior–posterior direction from the incisors (Y-coordinate) and also transversely from left to right from the mid-sagittal plane (X-coordinate).
      The activity of masticatory muscles, measured with electromyography, and CFs, measured with a digital occlusal analyzer, have been studied together.
      • Li B.Y.
      • Zhou L.J.
      • Guo S.X.
      • Zhang Y.
      • Lu L.
      • Wang M.Q.
      An investigation on the simultaneously recorded occlusion contact and surface electromyographic activity for patients with unilateral temporomandibular disorders pain.
      ,
      • Prabahar T.
      • Gupta N.
      • Chowdhary N.
      • Sonnahalli N.K.
      • Chowdhary R.
      • Reddy V.R.
      Comparative evaluation of occlusal bite force in relation to the muscle activity in the mixed dentition children of age group 9–12 years: a t-scan analysis.
      ,
      • Wieczorek A.
      • Loster J.
      • Loster B.W.
      Relationship between occlusal force distribution and the activity of masseter and anterior temporalis muscles in asymptomatic young adults.
      However, the authors are unaware of in vitro studies evaluating the consequence of the lack of masticatory force symmetry.
      The purpose of this study was to analyze whether articulated dental casts have a unique resulting CF at MIP, which could only be changed by modifying the geometry of the dentition, or, alternatively, if there is an influence of CO-MF even if the magnitude of the masticatory forces are kept constant. The null hypotheses were that, for any articulated dental cast, no relationship would be found between CO-MF and CO-CF when the magnitude of the masticatory forces was constant and that, for any articulated dental cast, no relationship would be found between CO-MF and the contact force magnitude (CF-M) when the magnitude of the masticatory forces was constant.

      Material and Methods

      The influence of the CO-MF position in the location of the CO-CF and the CF-M was studied by applying an input force (equivalent to masticatory force) of constant magnitude at 9 different positions on articulated dental casts (equivalent to different CO-MF or balances of masticatory force). Once approval had been obtained from the university ethical committee (M10_2019_254), 3 different dental casts (CASE GSR, CASE GSL, and CASE BSL) were mounted in maximum intercuspation on a semiadjustable articulator (Artex CN; Amann Girrbach AG). The articulator was calibrated by using a magnetic plate system, which obtained a precision where deviations were reduced to below 10 μm. An investigator made an impression of both arches (3M ESPE Express 2 Putty Soft and 3M Express 2 Light Body Standard; 3M), recorded the occlusion at MIP with silicone (3M Imprint 4 Bite; 3M), and determined the position of the maxilla and its orientation relative to the cranium axis by using a facebow (Artex facebow; Amann Girrbach AG). That position was transferred to the articulator in the laboratory by using a transfer table (Artex Transfer Table; Amann Girrbach AG). Dental casts were mounted using the transferred position and the occlusal registry at MIP. This standardized procedure ensured the correct positioning of the dental cast in relation to the Frankfurt plane on the semiadjustable articulator.
      The dental casts were selected under dental criteria: CASE GSR (good symmetry and right displaced) presented many contact points, a good left–right symmetry in its contact points, and a CO-CF slightly shifted to the right; CASE GSL (good symmetry and left displaced) also presented many symmetrical contact points, but its CO-CF was slightly shifted to the left; and finally, CASE BSL (bad symmetry and left displaced) presented few nonsymmetrical contact points, and its CO-CF was clearly displaced to the left from the mid-sagittal plane.
      A pattern with an XY coordinate system (X=transverse axis and Y=longitudinal axis) and 9 points for application of input force was attached to the upper arm of the semiadjustable articulator (Fig. 3). The points for application of the input force were established by using a grid with 3 rows and 3 columns: an anterior row, placed 6-mm distally from the incisors; a middle row, at a distance of 28 mm from the incisors; a posterior row, at a distance of 50 mm; a first column, placed 18 mm to the right from the mid-sagittal plane; a second column exactly in the mid-sagittal plane; and a third column, placed 18 mm to the left from the mid-sagittal plane. Furthermore, the 9 points for application of input force were grouped in 3 positional groups according to the row in which they were situated: anterior positions, the 3 situated in the anterior row; middle points, the 3 situated in the middle row; and posterior positions, the 3 situated in the posterior row.
      Figure thumbnail gr3
      Figure 3Pattern to control position of input force application point.
      For each articulator mounting, a digital occlusal analyzer (T-Scan; Tekscan Inc) was positioned with the aid of a fixed structure that had been custom designed and manufactured (Fig. 4). The structure was then placed on the table of a vertical drill (OPTI F30; Optimum), and an input force equivalent to a masticatory force of constant magnitude of 169 N—previously measured with a digital dynamometer (Beslands Push-pull Force Gauge, SF-500; beslandstool)—was applied perpendicular to the Frankfurt plane with a Ø12.5-mm spherical tip 10 times on each application point identified in the pattern (Fig. 3).
      Figure thumbnail gr4
      Figure 4Assembly of elements used for tests.
      Each time an input force was applied, the occlusion was measured with the digital occlusal analyzer, as per the manufacturer’s protocol.
      • Kerstein R.B.
      • Radke J.
      In-vitro consistency testing of the T-Scan 10 relative force measurement system in-vitro consistency testing of the T-Scan 10 relative force measurement system.
      All measurements were recorded in the software program of the digital occlusal analyzer, and, from each of them, 3 values were subtracted from the American Standard Code for Information Interchange (ASCII) format file: CF-M and 2 coordinates of CO-CF (X and Y, in millimeters and with the same reference system as the pattern). The X-coordinate corresponded to the transverse location of the CO-CF from the mid-sagittal plane, and the Y-coordinate to the longitudinal location in the anterior–posterior direction measured from the incisors. In total, in each mounting, 10 CF-M and X and Y coordinates of CO-CF were obtained for each input force application point.
      All measurements were entered into a statistical software program (IBM SPSS Statistics, v26; IBM Corp) to test the null hypotheses. To test the first null hypothesis, the influence of the input force application position in the X and Y coordinates of the CO-CF was analyzed with a 2-way repeated measures ANOVA test (α=.05) for each of the coordinates. Each coordinate was considered a repeated measures factor with 9 levels, and the articulated dental casts an independent factor with 3 levels (given that 10 repeated coordinates were measured for each of the 9 force application positions for each of the 3 articulated dental casts). Similarly, to test the second null hypothesis, the influence of the input force application position on the CF-M was analyzed by comparing the magnitudes for each group of input force application positions (anterior, middle, and posterior). In addition, to perform the comparisons, the means of the 10 CF-Ms measured for each of the 9 input force application positions were calculated for each of the articulated dental casts. Then, the CF-Ms measured for each of the articulated dental casts were divided by the highest mean value and expressed as percentages. A 2-way repeated measures ANOVA test (α=.05) was then performed, considering CF-M (expressed in percentages) a repeated measures factor with 3 levels and the articulated dental casts an independent factor with 3 levels (given that 10 repeated CF-M values were obtained for each of the 3 input force application positions belonging to each of the 3 groups of input force application positions for each of the 3 articulated dental casts).

      Results

      The influence of the input force application point on the X and Y coordinates of CO-CF is presented in Figure 5. The 2-way repeated measures ANOVA test revealed significant differences in the X (FG-G [9.32, 125.85]=1204.70; P<.001; η2=0.989) and Y (FG-G [7.75, 104.66]=1149.69; P<.001; η2=0.988) coordinates among the groups for the 3 articulated dental casts, demonstrating that the point of application of the input force had an influence on their value. Furthermore, a coefficient of variation mean of 0.37% was obtained in the X and Y coordinate values obtained in the 10 repetitions for the 27 groups. In Figure 6, the median of the CO-CF location for each group of repetitions of the test variable CO-MF has been positioned with a symbol; that is, 9 symbols for each dental cast. A reference ellipse has also been drawn, within which is what is considered to be a normal location in MIP for 68% of the population, according to Maness et al
      • Maness W.L.
      • Podoloff R.
      Distribution of occlusal contacts in maximum intercuspation.
      and Mizui et al.
      • Mizui M.
      • Nabeshima F.
      • Tosa J.
      • Tanaka M.
      • Kawazoe T.
      Quantitative analysis of occlusal balance in intercuspal position using the T-Scan system.
      Figure thumbnail gr5
      Figure 5Box plot for XY coordinates for location of CO-CF. A, X-coordinate for CASE GSR (yellow). B, Y-coordinate for CASE GSR. C, X-coordinate for CASE GSL (gray). D, Y-coordinate CASE GSL. E, X-coordinate for CASE BSL (blue). F, Y-coordinate for CASE BSL. BSL, bad symmetry and left displaced; CO-CF, center of contact force; GSL, good symmetry and left displaced; GSR, good symmetry and right displaced.
      Figure thumbnail gr6
      Figure 6Graph for medians of locations for CO-CF in 3 dental cast situations. Each symbol corresponds to a different position of CO-MF. CO-CF, center of contact force; CO-MF, center of masticatory forces.
      Regarding the influence of the input force application point in the magnitude data of CF (Fig. 7), a 2-way repeated measures ANOVA test (FG-G [3.37, 146.71]=73.08; P<.001; η2=0.627) revealed statistically significant differences in CF magnitude percentages among the 3 groups in each articulated dental cast, demonstrating that even when testing with the same magnitude of input force (169 N), the values of CF-M varied significantly for the 3 groups. In the CF-M percentages, a coefficient of variation mean of 3.70% was obtained for 30 repetitions for the 3 groups in each of the 3 articulated dental casts.
      Figure thumbnail gr7
      Figure 7Box plot of CF magnitudes percentages for 3 dental cast cases. Each group corresponds to anterior, middle, and posterior positions of CO-MF. BSL, bad symmetry and left displaced; CF, contact force; CO-MF, center of masticatory forces; GSL, good symmetry and left displaced; GSR, good symmetry and right displaced.

      Discussion

      The purpose of this study was to analyze whether a change in the masticatory force balance influenced the location of the CO-CF and its CF-M. Statistically significant differences were found on the measured CO-CF and CF-M values when changing the masticatory force balances (application points). Therefore, the hypothesis that each articulated dental cast had a unique location of the CO-CF at the MIP was rejected as was the hypothesis that if the intensity of the masticatory force remained constant, the magnitude of the tooth CF would also always remain constant. Furthermore, the obtained average coefficient of variation (0.37% in location of CO-CF and 3.70% in CF-M) revealed that the test method designed and used to measure both parameters of dental CFs was highly repeatable.
      To analyze the location of CO-MF, 27 groups (9 positions of CO-MF in 3 dentures) were used, with excellent repeatability. However, to analyze the magnitude of CFs, 9 groups (anterior, middle, and posterior positions of CO-MF in 3 dentures) were used; these were less homogeneous and therefore had reduced repeatability.
      Cerna et al
      • Cerna M.
      • Ferreira R.
      • Zaror C.
      • Navarro P.
      • Sandoval P.
      Validity and reliability of the T-Scan III for measuring force under laboratory conditions.
      concluded that the digital occlusal analyzer used in the present study was a reliable device if the total force value was measured with sensors of the same manufacturing series, as used in the present study. Kerstein et al
      • Kerstein R.B.
      • Radke J.
      In-vitro consistency testing of the T-Scan 10 relative force measurement system in-vitro consistency testing of the T-Scan 10 relative force measurement system.
      also concluded that this digital occlusal analyzer was a reliable device if the percentage of left and right side force were measured; this information is consistent with good repeatability of the X coordinate of the CO-CF in the present study.
      Lee et al
      • Lee W.
      • Kwon H.B.
      • Kim M.J.
      • Lim Y.J.
      Determination of the reliability and repeatability of a quantitative occlusal analyzer by using a piezoelectric film sensor: an in vitro study.
      also concluded that the digital occlusal analyzer was a reliable device if absolute force value is measured. In their in vitro test, a denture was mounted in a semiadjustable articulator and was compressed with a force applied always at the same point of the articulator. This conclusion was consistent with the results of the present study (Fig. 6).
      ANOVA tests performed on 2 coordinates of the CO-CF for 3 dental modes led to the rejection of the first null hypothesis (P<.05). Graphically (Fig. 6), it could be clearly distinguished that medians of the CO-CF varied when the CO-MF changed. Examining the graph also led to the rejection of the first null hypothesis because each articulated dental cast did not have a unique location for the CO-CF in MIP. By using electromyography, Wiechens et al
      • Wiechens B.
      • Brockmeyer P.
      • Wassmann T.
      • Rödiger M.
      • Wiessner A.
      • Bürgers R.
      Time of day-dependent deviations in dynamic and static occlusion: a prospective clinical study.
      concluded that symmetrical masticatory forces do not imply a centered and balanced CO-CF, consistent with the results of the present study (Fig. 6).
      In the present study, the quality of the occlusion of articulated dental casts did not significantly influence variations in the location of the CO-CF. However, the poorest quality dental cast, CASE BSL, presented less variation in the CO-CF than better quality casts, possibly because there were fewer contact points and the redistribution of forces was more limited.
      Although longitudinal position changes made in the CO-MF were greater than transverse changes, variations in the longitudinal location of the CO-CF were smaller. Furthermore, the transverse location variation of the CO-CF directly correlated with the transverse position changes of CO-MF. However, the longitudinal CO-CF location variation did not directly correlate with the longitudinal CO-MF changes. This effect may have been because transverse stability was achieved with 2 supports on the condyles located at each transverse end, but longitudinal stability was only achieved with the 2 condyles located at the same longitudinal end.
      ANOVA tests for the percentages of the CF-M also rejected the second null hypothesis (P<.05). When the same value of masticatory force (value 169 N) was applied in positions closer to condyles, the magnitude of CFs was lower (Fig. 7), because condyles bear a larger part of the masticatory force.
      Among the studies simultaneously analyzing muscle force and dental CF,
      • Shimada A.
      • Baad-Hansen L.
      • Svensson P.
      Effect of experimental jaw muscle pain on dynamic bite force during mastication.
      • Goiato M.C.
      • Zuim P.R.J.
      • Moreno A.
      • et al.
      Does pain in the masseter and anterior temporal muscles influence maximal bite force?.
      • Prabahar T.
      • Gupta N.
      • Chowdhary N.
      • Sonnahalli N.K.
      • Chowdhary R.
      • Reddy V.R.
      Comparative evaluation of occlusal bite force in relation to the muscle activity in the mixed dentition children of age group 9–12 years: a t-scan analysis.
      • Wieczorek A.
      • Loster J.
      • Loster B.W.
      Relationship between occlusal force distribution and the activity of masseter and anterior temporalis muscles in asymptomatic young adults.
      none have focused on the influence of the first force with respect to the second. Wiechens et al
      • Wiechens B.
      • Brockmeyer P.
      • Wassmann T.
      • Rödiger M.
      • Wiessner A.
      • Bürgers R.
      Time of day-dependent deviations in dynamic and static occlusion: a prospective clinical study.
      and Berry and Singh
      • Berry D.C.
      • Singh B.P.
      Daily variations in occlusal contacts.
      concluded that the location and intensity of CF changes throughout the day and pointed to variation in muscle contraction as the likely cause, a finding that was consistent with the present study.
      Limitations of the present study include the use of in vitro tests. Future research could measure the variability in the CO-CF in vivo by monitoring muscle contraction with surface electromyography measurements. To do this, once the point of maximum intercuspation is reached, patients could voluntarily vary muscle contraction from left to right to determine the effect on the location of the CO-CF.
      Although an imbalance was introduced by shifting the CO-MF from the mid-sagittal plane, another limitation was the impossibility of quantifying muscle force on each side. In order to quantify it, it would be necessary to know the exact lateral position of muscles, and this aspect was beyond the scope of the present study.
      The null hypotheses were established for any existing dental cast, and they must be satisfied for all situations. Three dental casts were analyzed, and the statistics determined that the null hypotheses were not satisfied; as a result, they are not satisfied in any situation. It might be thought that the null hypotheses will not be satisfied in any dental cast, but, with only 3 examples tested, the statistical evidence was insufficient to confirm this.

      Conclusions

      Based on the findings of this in vitro study, the following conclusions were drawn:
      • 1.
        The location of the CFs offered by a digital occlusal analyzer at the maximum intercuspation position was not necessarily unique for each articulated dental cast. Even if the intensity of the masticatory force remains unchanged, changes in its lateral or longitudinal balance also influence this location.
      • 2.
        Even if the intensity of the masticatory force remains constant, the magnitude of the absolute force measured with a digital occlusal analyzer can vary significantly. Changes in the longitudinal position of the CO-MF cause this variation.

      CRediT authorship contribution statement

      Mikel Jauregi: Conceptualization, Methodology, Software, Investigation, Writing – original draft. Xabier Amezua: Investigation, Writing – original draft, Visualization. Angel P. Manso: Formal analysis, Data curation, Visualization. Eneko Solaberrieta: Validation, Writing – review & editing, Supervision, Project administration, Funding acquisition.

      Acknowledgments

      The authors thank the University of the Basque Country (UPV/EHU) for providing the DEHI research laboratory (www.ehu.eus/dehi). The authors also thank Ayesta dental practice and Odontolan dental laboratory for their assistance.

      References

        • Maness W.L.
        • Chapman R.J.
        • Dario L.D.
        Laboratory evaluation of a direct reading digital occlusal sensor.
        J Prosthet Dent. 1985; 54: 744
        • Bozhkova T.P.
        The .T-SCAN system in evaluating occlusal contacts.
        Folia Med (Plovdiv). 2016; 58: 122-130
        • Olivieri F.
        • Kang K.H.
        • Hirayama H.
        • Maness W.L.
        New method for analyzing complete denture occlusion using the center of force concept: a clinical report.
        J Prosthet Dent. 1998; 80: 519-523
        • Shinogaya T.
        • Tanaka Y.
        • Toda S.
        • Hayakawa I.
        A new approach to evaluating occlusal support by analyzing the center of the bite force.
        Clin Oral Investig. 2002; 6: 249-256
        • Ferrato G.
        Digital evaluation of occlusal forces: comparison between healthy subjects and TMD patients.
        Ann Stomatol. 2017; 8: 79-88
        • Li B.Y.
        • Zhou L.J.
        • Guo S.X.
        • Zhang Y.
        • Lu L.
        • Wang M.Q.
        An investigation on the simultaneously recorded occlusion contact and surface electromyographic activity for patients with unilateral temporomandibular disorders pain.
        J Electromyogr Kinesiol. 2016; 28: 199-207
        • Liu C.W.
        • Chang Y.M.
        • Shen Y.F.
        • Hong H.H.
        Using the T-scan III system to analyze occlusal function in mandibular reconstruction patients: a pilot study.
        Biomed J. 2015; 38: 52-57
        • Wieczorek A.
        • Loster J.E.
        Activity of the masticatory muscles and occlusal contacts in young adults with and without orthodontic treatment.
        BMC Oral Health. 2015; 15: 116
        • Bandodkar S.
        • Tripathi S.
        • Chand P.
        • et al.
        A study to evaluate psychological and occlusal parameters in bruxism.
        J Oral Biol Craniofacial Res. 2022; 12: 38-41
        • Sierpinska T.
        • Kuc J.
        • Golebiewska M.
        Assessment of masticatory muscle activity and occlusion time in patients with advanced tooth wear.
        Arch Oral Biol. 2015; 60: 1346-1355
        • Di Berardino F.
        • Filipponi E.
        • Schiappadori M.
        • Forti S.
        • Zanetti D.
        • Cesarani A.
        The occlusal imaging and analysis system by T-scan III in tinnitus patients.
        Biomed J. 2016; 39: 139-144
        • Bozhkova T.
        • Musurlieva N.
        • Slavchev D.
        • Dimitrova M.
        Occlusal indicators used in dental practice: a Survey Study.
        Biomed Res Int. 2021; 2021: 2177385
        • Saraçoǧlu A.
        • Özpinar B.
        In vivo and in vitro evaluation of occlusal indicator sensitivity.
        J Prosthet Dent. 2002; 88: 522-526
        • Bozhkova T.
        • Musurlieva N.
        • Slavchev D.
        Comparative study qualitative and quantitative techniques in the study of occlusion.
        Biomed Res Int. 2021; 2021: 1163874
        • Bostancıoğlu S.E.
        • Toğay A.
        • Tamam E.
        Comparison of two different digital occlusal analysis methods.
        Clin Oral Investig. 2022; 26: 2095-2109
        • Fraile C.
        • Ferreiroa A.
        • Solaberrieta GP E.
        Intraoral versus extraoral digital occlusal records: a pilot study.
        Int J Comput Dent. 2018; 21: 329-333
        • Solaberrieta E.
        • Etxaniz O.
        • Otegi J.R.
        • Brizuela A.
        • Pradies G.
        Customized procedure to display T-Scan occlusal contacts.
        J Prosthet Dent. 2017; 117: 18-21
        • Cerna M.
        • Ferreira R.
        • Zaror C.
        • Navarro P.
        • Sandoval P.
        Validity and reliability of the T-Scan III for measuring force under laboratory conditions.
        J Oral Rehabil. 2015; 42: 544-551
        • Lee W.
        • Kwon H.B.
        • Kim M.J.
        • Lim Y.J.
        Determination of the reliability and repeatability of a quantitative occlusal analyzer by using a piezoelectric film sensor: an in vitro study.
        J Prosthet Dent. 2022; 127: 331-337
        • Kerstein R.B.
        • Radke J.
        In-vitro consistency testing of the T-Scan 10 relative force measurement system in-vitro consistency testing of the T-Scan 10 relative force measurement system.
        Adv Dent Technol Tech. 2022; 4: 47-58
        • Maness W.L.
        • Podoloff R.
        Distribution of occlusal contacts in maximum intercuspation.
        J Prosthet Dent. 1989; 62: 238-242
        • Mizui M.
        • Nabeshima F.
        • Tosa J.
        • Tanaka M.
        • Kawazoe T.
        Quantitative analysis of occlusal balance in intercuspal position using the T-Scan system.
        Int J Prosthodont. 1994; 7: 62-71
        • Türp J.C.
        • Greene C.S.
        • Strub J.R.
        Dental occlusion: a critical reflection on past, present and future concepts.
        J Oral Rehabil. 2008; 35: 446-453
        • Wiechens B.
        • Brockmeyer P.
        • Wassmann T.
        • Rödiger M.
        • Wiessner A.
        • Bürgers R.
        Time of day-dependent deviations in dynamic and static occlusion: a prospective clinical study.
        J Prosthet Dent. 7 July 2022; ([Epub ahead of print.])https://doi.org/10.1016/j.prosdent.2022.05.025
        • Berry D.C.
        • Singh B.P.
        Daily variations in occlusal contacts.
        J Prosthet Dent. 1983; 50: 386-391
        • Raadsheer M.C.
        • Van Eijden T.M.G.J.
        • Van Ginkel F.C.
        • Prahl-Andersen B.
        Contribution of jaw muscle size and craniofacial morphology to human bite force magnitude.
        J Dent Res. 1999; 78: 31-42
        • Peck C.C.
        Biomechanics of occlusion - implications for oral rehabilitation.
        J Oral Rehabil. 2016; 43: 205-214
        • Shimada A.
        • Baad-Hansen L.
        • Svensson P.
        Effect of experimental jaw muscle pain on dynamic bite force during mastication.
        Arch Oral Biol. 2015; 60: 256-266
        • Goiato M.C.
        • Zuim P.R.J.
        • Moreno A.
        • et al.
        Does pain in the masseter and anterior temporal muscles influence maximal bite force?.
        Arch Oral Biol. 2017; 83: 1-6
        • Prabahar T.
        • Gupta N.
        • Chowdhary N.
        • Sonnahalli N.K.
        • Chowdhary R.
        • Reddy V.R.
        Comparative evaluation of occlusal bite force in relation to the muscle activity in the mixed dentition children of age group 9–12 years: a t-scan analysis.
        Int J Clin Pediatr Dent. 2021; 14: S29-34
        • Wieczorek A.
        • Loster J.
        • Loster B.W.
        Relationship between occlusal force distribution and the activity of masseter and anterior temporalis muscles in asymptomatic young adults.
        Biomed Res Int. 2013; 2013: 354017